background image

10

Molift QuickRaiser 205

Safety devices

Molift Quick Raiser 205 is equipped with several 

safety devices, which are intended to prevent 

injury to personnel or damage to the equipment in 

the case of incorrect use or failure

The lifter has an overload sensor preventing the 

lifter to be operated if the load exceeds SWL.

The suspension is hinged to prevent the arm and 

suspension from squeezing the user.

Steering handles

Use the steering handles to move the lifter.

Do not push or pull the user, 

lifting column or lifting arm.  

This may cause stability prob-

lems and lifter may tip over.

The lifter shall only be used for transfer

over short distances. It is not a replacement

for wheelchair or similar.

Brakes

Rear wheels/castors are equipped with kick brakes. 

Step on the break to lock brakes. Push up or kick 

on top part of the brake to unlock.

Emergency stop / Emergency lowering

Emergency Stop

The emergency stop shuts off the power when 

pushed in. Turn clockwise to reset.

Electrical emergency lowering

Use a pen or sharp object to push and hold 

button for electrical emergency lowering to lower 

suspension. The electrical emergency lowering 

function does not work when the emergency stop 

is activated. 

Only use the electrical emergency 

lowering if no other button works.

Manual Emergency lowering

Use the wrench located in the bottom cover to flip 

the control panel cover off at the two points. 

Push in the emergency stop but-

ton before removing the cover. 

Rotating parts represents a squeez-

ing hazard!

Use the 17mm spanner and turn ball screw coun-

ter clockwise until the user is safely lowered. The 

manual emergency lowering function should only 

be used when the electrical emergency lowering 

does not work.

If emergency lowering is used, 

mark the lifter “out of order” and 

order service!

Содержание Quick Raiser 205

Страница 1: ...Molift Quick Raiser 205 EN User manual BM13001 Rev C 2018 01 08...

Страница 2: ...leg support 13 Using a 4 point lifting arm 14 Using a 2 point lifting arm 14 Raising 15 Transfer 15 Accessories 16 Accessories 16 Suspensions and Slings 16 Maintenance 17 Cleaning and disinfeciton 17...

Страница 3: ...rs and an electric motor powered by a battery The hoist is operated by using the hand control or the control panel on the hoist It has electrically adjustable legs and with a lifting capacity of 205 k...

Страница 4: ...es class 1 and has been tested and approved by a third party according to standards IEC 60601 1 IEC 60601 1 2 and NS EN ISO 10535 2006 Conditions for Use Lift and transfer of a person will always pose...

Страница 5: ...Applied parts Sling Foot rest Rocker arm handles Technical Data Quick Raiser 205 922 mm 3619 64in 808 mm 3157 64in 574 mm 2219 32in 530 mm 2055 64in 1068 mm 423 64in 1320 mm 5131 32in 1730 mm 687 64i...

Страница 6: ...All screws with pre applied threadlock should only be used once Spare part art no 2920075 1 Battery Push in the emergency stop button before assembly Place the column on a horizontal surface Make sure...

Страница 7: ...sis Mount the screw with washer on column Slide the screw down to chassis and tighten the screw Turn the lifter over on the side Insert screw and washer in the top hole and tighten the screw Connect c...

Страница 8: ...ert the bolt and tighten the screw using the 17mm spanner and an 6mm allen key Make sure the lifting arm is properly fastened to the trolley on the column Fit the cover onto the trolley 5 Mains cable...

Страница 9: ...ns Make sure the leg support and lifting arm are properly fixed and that there are no loose parts on the lifter Make sure that the lifting column is properly fixed to the chassis with both screws tigh...

Страница 10: ...inimum 18 minutes Duty Cycle 10 Intermittence according to standard ISO EN 10535 Safe Working Load The lifter is marked with Safe Working Load SWL If maximum load SWL differs between lifter suspension...

Страница 11: ...ush up or kick on top part of the brake to unlock Emergency stop Emergency lowering Emergency Stop The emergency stop shuts off the power when pushed in Turn clockwise to reset Electrical emergency lo...

Страница 12: ...24V SLA battery or a 25 6V LiFePO4 battery both with an integrated charger The SLA battery has a life expectancy of approx 500 charge cycles The LiFePO4 battery has a life expectancy of approx 2000 c...

Страница 13: ...odic inspection required Red Perform service Charging LED The lifter has an integrated charger Park the lifter and connect to a wall socket to charge the battery A LED light on the back of the battery...

Страница 14: ...turers Adjusting leg support Move the Quick Raiser 205 in front of the user Place the feet on the footplate If necessary help the user to place the feet correctly Lock the wheels on the lifter before...

Страница 15: ...lightly when lifting and if possible hold onto the handles Removing sling Loosen the straps again by angling the buckle and puling the strap out Using a 2 point suspension On the 2 point suspension ar...

Страница 16: ...y fastened 5 Lift user to standing postition and tighten and adjust the sling if necessary Make sure the users hands are placed correctly to avoid that the lifting straps are squeezing the users hands...

Страница 17: ...ns Art no 2910059 Set of small cushions Art no 2910076 Lifting arms and slings 4 point suspension Quick Raiser Active lifting arm Art no 2910060 Rgosling Active Art no 1720910 Size XS Art no 1720920 S...

Страница 18: ...able on request Periodic Inspection Scope Periodic inspection is a visual examination particu larly of the lifter s load bearing structure and lifting mechanism with attachments brakes controls safety...

Страница 19: ...Short circuit between the hand control and lifter clean the connectors of the hand control and chassis with spirits to remove grease and dirt The hand control has failed exchange hand control contact...

Страница 20: ......

Отзывы: