SpectraMax MiniMax Imaging Cytometer Installation Guide
24
5024111 D
5. Push the two retaining clips on either side of the detection cartridge so that they fasten
securely to the retaining rods on both sides of the detection cartridge drawer.
Note:
Detection cartridges that occupy more than one slot have two retaining
clips on each side. Securely fasten all retaining clips to the retaining rods on both
sides of the detection cartridge drawer.
Item
Detection Cartridge Retaining Clips and Retaining Rods
1
Detection cartridge
2
Retaining clip unattached
3
Retaining clip attached
4
Retaining rod
5
Retaining rod
6. Touch
to close the detection cartridge drawer.
7. Start the SoftMax Pro Software and connect to the instrument.
If the software is running and connected to the instrument, select the Operations tab
and click
Refresh
to let the software detect the detection cartridges.