Robotic Integration
264
0112-0109 H
waste_bottleb<SP>ok<TAB>
fill_bottlea<SP>ok<TAB>
fill_bottleb<SP>ok<TAB>
tips<SP>on<TAB>
outer_auto_door<SP>open<TAB>
tip_washer<SP>ok<TAB>
cell_reservoir<SP>not_ok,0<TAB>
chiller_status<SP>ok,18.0, 18.0<TAB>
intensifier_status<SP>ok<CR>
Cellflaskcontrol<TAB>Rate<CR>
Command Description
This command sets the stir-rate for the cell-flask control. After sending
this command, the robotic controller should monitor the instrument
stir-rate to confirm that the instrument reaches the setpoint.
Command Parameters
There is one required parameter, rate, for this command.
If a value of zero is used for the rate parameter, the cell flask control
will stop stirring. If a valid value for rate is used, the cell flask control
will be turned on and stir at this rate. Valid values for rate are integers
between 1–25.
Example Command Line
cellflaskcontrol<TAB>22<CR>
Example Response
ok<CR>
Suggested Steps in Command usage
•
Make sure an experiment is not running.
•
Send
cellflaskcontrol
command.
•
Response will be OK, confirming that command execution has
begun, or a Condition code related to the command indicating
that execution could not begin.
•
Monitor status until the instrument reports that it has reached
the requested stir-rate.
Command Specific Error Responses
C76–Cell flask stirring rate parameter was not sent
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...