FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
219
Troubleshooting Robotic Integration
Data parameters need to
be reset (subtract bias,
spatial uniformity
correction, positive
control,
etc
.) for each
experiment.
*.vamp file
parameters not set up
correctly.
Select the
Grouping
or
Correction
buttons to
adjust the parameters as
necessary.
False negative wells.
Pipettor adjustment
problem.
Tip problem.
Contact Technical Support
at
+1-800-635-5577.
Table 9-9
Troubleshooting Data (cont’d)
Symptom or Error
Message
Possible Causes
Solutions
Table 9-10
Troubleshooting Robotic Integration
Symptom or Error
Message
Possible Causes
Solutions
220: “Unable to load plate. No
plate on landing pad.”
Plate present on
plate landing pad,
but plate presence
was not detected.
Contact Technical
Support at
+1-800-635-5577.
If plate not
positioned properly
by robot, contact
Technical Support for
robot manufacturer.
No plate was
delivered to landing
pad by 3
rd
party
robot.
Contact Technical
Support for robot
manufacturer.
221: “Unable to load plate. Plate
already detected at requested
location.”
Plate already at
requested position.
Unload plate
currently in the
desired position prior
to delivering new
plate.
If robot is attempting
to load multiple
plates to one position
without retrieving
plates, contact
Technical Support for
robot manufacturer.
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...