EMax® Plus Microplate Reader Quick Start Guide

2

5032019 B

Removing the Shipping Lock From the Microplate Carrier

A red metal lock secures the microplate
carrier during shipping. This lock must be
removed before the instrument can be
used.

To remove the shipping lock, place one
hand on top of the lock, and then use
your other hand to carefully pull the
bottom of the lock until it comes free of
the instrument.

Note:

Keep the shipping lock in

case of future transport needs.

Connecting the Instrument Cables

The power cord and USB cable connect to the ports on the rear of the instrument.

Connect the instrument to the controlling computer using the supplied USB cable. Use
the power cord supplied with the instrument to connect the instrument to a grounded
wall outlet.

To connect the instrument cables:

1. Make sure that the instrument and host computer are placed on a dry, flat work

area with sufficient space for both devices and the required cables. To ensure
sufficient ventilation and provide access for disconnecting power from the
instrument, maintain a 20 cm to 30 cm (7.9 in. to 11.8 in.) gap between the rear of
the instrument and the wall.

2. Make sure that the power switch on the rear of the instrument is in the

Off position.
Do not turn on the power to the instrument until after the software has been
installed. See the SoftMax Pro Software installation guide.

Note:

The instrument is supported by SoftMax Pro Software version 6.4.0.1

or later.

3. Connect one end of the supplied USB cable to one of the USB ports on the

computer, and then connect the other end of the USB cable to the USB port on the
rear of the instrument.

4. Connect the supplied power adapter to the power port on the rear of the

instrument, and then connect the other end of the power adapter to a grounded
electrical wall outlet.

Содержание EMax Plus

Страница 1: ...equirements for the SoftMax Pro Software See the SoftMax Pro Software installation guide Setting up the instrument includes the following procedures Installing the SoftMax Pro Software See the SoftMax Pro Software installation guide Removing the Shipping Lock From the Microplate Carrier see page 2 Connecting the Instrument Cables see page 2 Starting the Instrument and the Software see page 3 Selec...

Страница 2: ...1 Make sure that the instrument and host computer are placed on a dry flat work area with sufficient space for both devices and the required cables To ensure sufficient ventilation and provide access for disconnecting power from the instrument maintain a 20 cm to 30 cm 7 9 in to 11 8 in gap between the rear of the instrument and the wall 2 Make sure that the power switch on the rear of the instrum...

Страница 3: ... SoftMax Pro Software installation guide Molecular Devices recommends turning the power off when the instrument is not in use Activating Your Software License When you first start the SoftMax Pro Software the Software License Activation dialog is displayed If you are running a trial version of the software click the Help tab in the ribbon and then click Software License To activate your SoftMax Pr...

Страница 4: ...plication help or user guide The instrument icon shows the connection status of the instrument Icon Status Instrument Connected Instrument Disconnected Instrument Simulated Table Q 1 Instrument Connection Status If the instrument status shows as disconnected check to make sure that the instrument is powered on and that the connections between the instrument and the computer are secure To select th...

Страница 5: ...s on page 5 Setting Up an Absorbance Endpoint Experiment on page 7 Loading and Unloading Microplates The microplate carrier on the EMax Plus Microplate Reader moves in and out of the read chamber during the read and then returns to the home position after the read ends If the microplate carrier does not return to the home position you can initialize the carrier to return it to its home position by...

Страница 6: ...rument can read 96 well clear bottom microplates and strip wells For most applications use microplates with flat bottom or round bottom wells Molecular Devices recommends that you do not use V bottom microplates unless an application specifically requires their use Irregular plastic density in the tip of the well can cause inaccurate measurements The type of microplate and the way it is handled ca...

Страница 7: ...lect a contiguous rectangular region on the microplate drag the cursor to select the wells to be read Columns do not need to start with 1 but must be contiguous To read all the wells in the microplate select the Select All check box Note If you select a subset of wells the instrument reads the entire microplate but reports the data for the selected wells only 8 Click OK 9 Save the settings as a da...

Страница 8: ...resentatives You can contact your local representative or contact Molecular Devices Technical Support by telephone at 800 635 5577 U S only or 1 408 747 1700 In Europe call 44 0 118 944 8000 Please have the system ID number system serial number software version number and the system owner s name available when you call For research use only Not for use in diagnostic procedures The trademarks menti...

Отзывы: