4. Technical Information

Technical specifications

Electrical specifications

Power supply

Input:

100 - 240 VAC, 50 - 60 Hz, 0.4 A

Output: 5 V - 2 A, max. 10 W

Battery

Type:

Lithium-Ion (LiOn)
5200 mAh

Expected life

1

:

Approx.  1½  year

Measurements per charge: 

Approx. 2000

Charging time:

Max. 5 hours

1

 At 10 hours daily use (no Wi-Fi activity)

Connectivity

USB

Micro-B USB (2.0)

(For charging only)

Wi-Fi

b/g/n, “Create Network” or “Connect to Network” Mode
WPA2, WPA2 Enterprise

Mechanical specifications

Analyzer

Size:

75 x 100 x 175 mm (H x W x D)

Weight:

Approx. 0.7 kg

Box of one analyzer

Size:

295 x 135 x 355 mm (H x W x D)

Weight:

Approx. 2.1 kg

IP Class

IP20

Ambient temperature

Operation:

0 to 40 °C, < 95% RH, non condensing

Storage: 

- 20 to 60 °C, < 95% RH, non condensing

Ambient pressure

Operation:

800 to 1100 mbar

Gas inlet

Sample hose L=600 mm, Ø0.5 mm

Содержание CheckPoint 3

Страница 1: ...P N 340486 B 09 2017 CheckPoint3 UserGuide EN...

Страница 2: ...This blank page has been inserted to enable double sided printing of the document...

Страница 3: ...NSENSOR CheckPoint 3 EN User Guide P N 340486 B 09 2017 CheckPoint3 UserGuide Published by MOCON Europe A S R nnedevej 18 4100 Ringsted Denmark Tel 45 57 66 00 88 info eu mocon com www dansensor com E...

Страница 4: ...DANSENSOR 2 User Guide EN CheckPoint 3 P N 340486 B 09 2017...

Страница 5: ...rview 10 Accessories 11 Setting up 12 Assembling the sample gas hose 12 Fitting the bumpers option 12 2 Operation and Maintenance 13 Basics 13 Switching on off 13 Restarting resetting 13 Touch screen...

Страница 6: ...52 Diagnostics 54 Setup 56 Data log 57 Network 58 Users 59 Date Time Units 61 Power save 62 Supervisor PIN code 63 Display 64 Access level 65 Calibration 66 Calibrate 0 CO2 67 Calibrate 100 CO2 68 Res...

Страница 7: ...74 Options 74 Appendix 75 Safety and handling instructions 75 Handling and maintenance 75 Repairing 75 Needles 75 Battery and charging 76 Prolonged heat exposure 76 Explosive atmospheres 76 Cleaning...

Страница 8: ...DANSENSOR 6 User Guide EN CheckPoint 3 P N 340486 B 09 2017...

Страница 9: ...e to CheckPoint 3 or other property Prior to using the equipment it is assumed that it has been properly installed and configured as described in this manual Always refer to this manual before operati...

Страница 10: ...and or act according to the information in order to obtain the best possible function of the equipment CAUTION The operator must observe and or act according to the information in order to avoid any...

Страница 11: ...rol of several CheckPoint 3 units in a production environment All measurement data can be saved in the device s database and exported for further analysis Models 3 different models are available Check...

Страница 12: ...placed in the retainer after measuring 3 5 colour touch display For intuitive operation of the device by use of explanatory icons and easy understandable text messages and buttons The Button The butto...

Страница 13: ...ing the battery on page 20 for details Septa 100 pcs When performing a measurement a septum should be applied to the package in order to ensure a leak free gas extraction and to protect the package fr...

Страница 14: ...dle in the retainer WARNING Due to safety considerations the needle should always be placed in the retainer after measuring Fitting the bumpers option 1 Fit the bumpers onto the side plates starting f...

Страница 15: ...wn Restarting resetting If device freezes or something is not working right try restarting the device by pressing and holding the button for approx 8 s Touch screen Except for the button on top of the...

Страница 16: ...s and their functions are listed below Button type Function Opens a submenu Executes function Returns to the previous window Undo editing end return to previous window When editing a product tapping t...

Страница 17: ...ng only offers a numerical keyboard All typing keys except for the number keys holds 2 or more symbols or characters Tap the respective key repeatedly until the required character appears The key dele...

Страница 18: ...ls field is empty Icon for Extended Data Pack Premium models only Software version Service Point Tapping this icon will display the company information for your appointed service provider Tap button t...

Страница 19: ...ull access you have to change the device to Supervisor access level To do so select Access level from the main Menu see Access level on page 65 for details No login default setting At start up the dev...

Страница 20: ...s Time Date Current time and date Measuring result Result of the recent measurement Only values selected for display in the product setup are displayed see New product on page 49 for details Black tex...

Страница 21: ...s a button giving fast access to the Wi Fi on off setting In this case the operator can turn the Wi Fi module on off without being logged in as Supervisor See Network on page 58 for details Measuremen...

Страница 22: ...when in power save mode simply touch the screen or press the button Power save settings are described in Power save on page 62 Charging and monitoring the battery WARNING For important safety informa...

Страница 23: ...the power supply to the main power outlet 4 When charging if the device is switched on and showing the measuring screen the battery indicator in the upper right corner of the display will show the cha...

Страница 24: ...tapping Yes or NO to decide how to proceed Error messages are mainly related to the device itself The Error message screen proposes the appropriate remedy to rectify the problem and must be acknowledg...

Страница 25: ...en as not acceptable see Alarm on page 46 for details Premium models Gas concentration alarms are set individually for each product see Products on page 47 for details NOTE You may experience shorter...

Страница 26: ...uct at the top in orange 2 Use and keys to scroll through the list to find the appropriate product NOTE If you have a large number of products scrolling through the list may be quite time consuming In...

Страница 27: ...e the septum with the needle so that the gas can be sucked in from the package through the tip CAUTION Make sure that the needle does not touch the product fluid or anything else in the package as thi...

Страница 28: ...and the icon changes to Please note that you have to do this within 30 s after the measurement has been performed otherwise the measurement will automatically be marked as valid You are also able to...

Страница 29: ...a administration much easier As on the device the WebGUI allows you to e g create and edit products and users and to view measurement data and statistics but in addition it also makes it possible to e...

Страница 30: ...tion on the device offers 2 ways to establish the network connection between CheckPoint 3 and the computer tablet Connect to Network Create Network Connect to Network Requirements An active Access poi...

Страница 31: ...abled default or Disabled NOTE Before selecting the Disabled setting you should consult your network administrator as incorrect settings can result in reduced or no network activity 8 Return to the pr...

Страница 32: ...let 1 Open a web browser 2 Type the IP address as acquired earlier e g 172 25 4 106 in the browser address bar and press Enter to connect and start the WebGUI application 3 Save the connection as a Fa...

Страница 33: ...WPA2 PSK 8 Enter Password if Security is set to WPA2 PSK 9 Return to Network screen and switch Wi Fi module on wait a few seconds 10 Select Info Status 11 Confirm that State is Working 12 Write down I...

Страница 34: ...but the network connection is maintained In general the WebGUI functions work like the corresponding functions on the device and reference is therefore made to the descriptions of these functions The...

Страница 35: ...2017 Home The Home screen shows the following information about the connected device If the below message appears you should go to the Date Time Units tab on the Setup screen to synchronize the devic...

Страница 36: ...o deselect it Select a product to Edit Copy or Delete it The only function that works with multiple selected products is Delete Find product name s product ID Enter t in the respective search field to...

Страница 37: ...lect multiple products Select a product to Delete data or to Export data collections for this product Export and import of data is described in Export Import on page 40 The Collected data tab shows th...

Страница 38: ...tab When scrolling up and down through the list on the Collected data tab and then changing back to the Collected data tab the entry currently on top on the Collected data tab will be the rightmost en...

Страница 39: ...DANSENSOR CheckPoint 3 EN User Guide 37 P N 340486 B 09 2017 Diagnostics To use the Start test measurement function from here the Test measurement must also be selected on the device...

Страница 40: ...the device Users Shows User list in Last used order no sorting possible Date Time Units Check CheckPoint 3 time against PC time when connected If difference exceeds value as set in Notify if differen...

Страница 41: ...DANSENSOR CheckPoint 3 EN User Guide 39 P N 340486 B 09 2017 Access It is possible to have Supervisor access in the WebGUI while having only User access on the device and vice versa...

Страница 42: ...are stored as well as their filenames When using the Export functions the application proposes a file name containing information about the content the date and time and the device serial no content...

Страница 43: ...DANSENSOR CheckPoint 3 EN User Guide 41 P N 340486 B 09 2017 Language When selecting a language all text throughout the WebGUI will be displayed in this language...

Страница 44: ...oft lint free cloth with a mild soap solution Avoid getting moisture in openings Replacing filter needle and measuring gas hose During use of the device dust will over time be sucked through the measu...

Страница 45: ...recommend that you perform the test as described below Check hose tightness 1 Remove the needle but keep the filter in place 2 Go to Diagnostics Test measurement 3 Read the values for P1 pressure and...

Страница 46: ...DANSENSOR 44 User Guide EN CheckPoint 3 P N 340486 B 09 2017...

Страница 47: ...page 47 for details Data log Premium models only View logged data for a product See Data log on page 51 for details Diagnostics Perform test measurements and view device serial nos SW version and coun...

Страница 48: ...appropriate value for the O2 concentration 0 1 100 Setting the alarm value to 0 will deactivate the alarm O2 Alarm type Select whether the O2 alarm should be Off an Upper alarm or a Lower alarm If ala...

Страница 49: ...wish to perform measurements on See Active Product on page 48 for details Edit Product Edit an existing product Product settings are described in New product on page 49 When having edited a product a...

Страница 50: ...it to select it and return to the measuring screen 3 If you have a large number of products scrolling through the list may be quite time consuming In this case use the Find function in the bottom of t...

Страница 51: ...Products menu Product name A unique product name up to 32 characters Product ID Unique product identification code up to 32 characters Measure CO2 Select whether or not the measured CO2 value should b...

Страница 52: ...uct screen 2 Select the appropriate product search method Selecting either the Name or Barcode methods brings up a touch screen keyboard for keying in the product data NOTE The Barcode search method w...

Страница 53: ...or the product The default selected product is always the active product product text is orange but any other product can be selected by using the Select Product function Select Product Select product...

Страница 54: ...ph shows layout when an upper limit 1 0 has been defined for the gas Graph shows layout when an upper limit 32 0 and a lower limit 27 0 has been defined for the gas Graph shows layout when no limits h...

Страница 55: ...ted When scrolling from page 1 to page 2 or 3 the rightmost entry from the graphs on page 1 will be shown on top on page 2 and 3 When scrolling up and down through the list on page 2 or 3 and then bac...

Страница 56: ...ut there will be no data logging Press button to start a test measurement When measurement runs values such as measured gas concentrations and pressures are continuously displayed The test measuring w...

Страница 57: ...7 About Displays a screen showing device information such as device serial no CPU ID no SW version and counters for next calibration measurements O2 sensor and pump Regulatory Displays a screen showin...

Страница 58: ...he Network setup screen See Network on page 58 for details Users Opens the Users setup screen See Users on page 59 for details Date Time Units Opens the Date Time Units setup screen See Date Time Unit...

Страница 59: ...buffer is full this warning appears If enabled When log buffer is full this warning appears In this case you must empty the log buffer using the Delete data or Delete all data function from the Data l...

Страница 60: ...nnection on page 28 for details Wi Fi setup Opens the setup screen for the currently selected network type See also Network connection on page 28 for details Wi Fi module ON Switches the Wi Fi module...

Страница 61: ...s or delete users The total number of users that can be created depends on the device model see Models on page 9 for details User list Displays a list of all users in alphabetical order The symbol ind...

Страница 62: ...Current user is the default User User login required At start up the device initializes and when finished you are taken to the User screen from where you must select a user to go to the measuring scre...

Страница 63: ...hh mm ss Date Setting of current date using Date format Date format Setting of date format DD MM YY or MM DD YY Time Date and Date format are related to the real time clock setting in the device The...

Страница 64: ...ter a measurement has been performed 0 999 s Dim display after Set the time of device inactivity after which the display switches into power save mode by reducing the display s back light intensity 10...

Страница 65: ...ed for changing from User access level to Supervisor access level To change the PIN code do the following 1 In the Setup menu select Supervisor PIN code This screen appears 2 Enter current PIN code an...

Страница 66: ...r e g Brightness brings up a window where value can be set from 1 to 10 by tapping the buttons Setting impact is seen while adjusting Restore defaults Resets the settings of the 3 display parameters b...

Страница 67: ...o select Access level from the Menu Tap Enter PIN code to open a touch screen keyboard and enter the pin code for Supervisor access level From the factory the pin code is set to 0000 The Supervisor pi...

Страница 68: ...For the 0 CO2 calibration 100 N2 technical air or atmospheric air can be used When using atmospheric air please ensure that there is no increased or unstable CO2 concentration inthe surroundingair as...

Страница 69: ...and then tap Yes 3 The pump starts to suck gas through the device and the current readout is displayed NOTE The pump will run for approx 40 s and if you do not tap Perform calibration within this tim...

Страница 70: ...irm and return to the Calibration start screen Calibrate 100 CO2 NOTE Before calibrating with 100 CO2 you must calibrate with 0 CO2 1 Select Calibrate 100 CO2 Now follow procedure as described for the...

Страница 71: ...N User Guide 69 P N 340486 B 09 2017 Language Selecting Language from the Menu will display a screen listing the available languages When selecting a language all text throughout the menus will be dis...

Страница 72: ...DANSENSOR 70 User Guide EN CheckPoint 3 P N 340486 B 09 2017...

Страница 73: ...urs 1 At 10 hours daily use no Wi Fi activity Connectivity USB Micro B USB 2 0 For charging only Wi Fi b g n Create Network or Connect to Network Mode WPA2 WPA2 Enterprise Mechanical specifications An...

Страница 74: ...r Sensor type Non Dispersive Infra Red NDIR single beam temperature compensated Range 0 100 Resolution 0 1 absolute Accuracy 2 0 absolute in range 0 to 20 3 0 absolute in range 20 to 100 Measuring tim...

Страница 75: ...ation conditions Ambient temperature 25 C Ambient pressure 1013 hPa Measurement gas temp 25 C Ambient relative humidity 40 RH Meas gas relative humidity 10 RH Measurement gas flow 45 ml min NOTE All g...

Страница 76: ...x Septum 15mm 1000 pcs grey 1 x Sampling kit CheckPoint 3 1 x Needle kit 0 8 x 16mm 10 pcs Cpl 1 x Filter sample gas 0 2 water trap 10 pcs Cpl Filter sample gas 0 2 water trap 10 pcs Ser Cpl P N 3103...

Страница 77: ...ade of metal glass and plastic and has sensitive electronic components inside The device can be damaged if dropped burned or crushed or if it comes in contact with liquid Do not use a damaged device s...

Страница 78: ...here your skin is in contact with the power supply when it is plugged into a power source for long periods of time For example do not place it under a blanket when it s plugged into a power source It...

Страница 79: ...tside of these temperature ranges Avoid exposing the device to dramatic changes in temperature or humidity WebGUI compatible browsers The WebGUI has been tested for compatibility with the following PC...

Страница 80: ...nted circuit board assembly Main PCB O O O O O O Pump O O O O O O Mounting hardware screws studs O O O O O O Internal cables O O O O O O Sample hose kit O O O O O O CheckPoint 3 Toxic or Hazardous Sub...

Страница 81: ...This blank page has been inserted to enable double sided printing of the document...

Страница 82: ...DANSENSOR MOCON Europe A S R nnedevej 18 4100 Ringsted Denmark Tel 45 57 66 00 88 info eu mocon com www dansensor com...

Отзывы: