22
User Guide
EN
Dansensor
®
Lippke
®
5000
Test head
The test head should be used together with the “Basic Stand” and the “Fixture for ASTM
restraining” measurement accessories.
The test head can be fitted with two different needles - see
for details.
The “Blind” plug
that comes with the test head can be used for testing if there are internal
leaks in the system.
To prepare the test head do the following:
1. Remove the “Blind” plug
, then fit the appropriate needle
or
on the test head.
The needles fit on the outside of the small needle
.
CAUTION! Needles should only be hand tightened - do not use any tools.
2. Connect air supply (Feed) and pressure measurement (Sense) hoses
to the device - see
for details.