52
User Guide
EN
Dansensor
®
Lippke
®
5000
Measurement accessory kits
The following measurement accessory kits can be used for a
Bubble
test:
Stand, basic
(with any test head and needle or tube adaptor, with or without filter)
Tube adapter
IV-Bag adapter
Hand-held, needle, mono
Hand-held needle, separate Feed and Sense
Bubble test settings
For remote controlled devices using the PC software, the test parameters are set in the PC
program.
For a stand-alone device the parameters are set on the device itself - see
for details.
A
Bubble
test is defined using the following parameters:
Parameters marked with an asterisk (*) are the only parameters that can be changed, when test
definition has already been used for testing and if the parameter
Settings ->
Device
->
Lock
Test
has been enabled.
Name
Product ID
Standard Test
*
State
Ventilation
*
Package Type
Package Connection
Package Volume (Vol.)
See explanation in
"Burst test settings" on page 40
Test Pressure (P)
Pressure at which the test is performed.
Test Time (T)
The period of time where the actual test is performed.
The test pressure is maintained.
Burst Detection Limit (L)
If a package for some reason expands suddenly during a
test, this will result in a pressure drop that could be
mistaken for a burst of the package.
Use this parameter to specify the max. accepted
pressure drop.
Test Speed (S)
If for some reason a package is difficult to fill, you can
lower the test speed to increase the fill time. Setting the
test speed to e.g. 50% will double the total fill time.
Custom Fields
*
Select if one or more of the 4 custom fields are required
for the current workflow - see
for details.
Содержание AMETEK Dansensor Lippke 5000
Страница 1: ...P N 380419 C 05 2022 UserGuide...
Страница 2: ...This blank page has been inserted to enable double sided printing of the document...
Страница 4: ...2 User Guide EN Dansensor Lippke 5000 P N 380419 C 05 2022...
Страница 38: ...36 User Guide EN Dansensor Lippke 5000 P N 380419 C 05 2022...
Страница 88: ...86 User Guide EN Dansensor Lippke 5000 P N 380419 C 05 2022...
Страница 106: ...104 User Guide EN Dansensor Lippke 5000 P N 380419 C 05 2022...
Страница 128: ...126 User Guide EN Dansensor Lippke 5000 P N 380419 C 05 2022...
Страница 135: ...This blank page has been inserted to enable double sided printing of the document...
Страница 136: ...0 0 VSPQF 4 3 OOFEFWFK 3JOHTUFE FONBSL 5FM JOGP EL NPDPO BNFUFL DPN XXX BNFUFLNPDPO DPN...