Note: This guide does not replace the user manuals.

Page 4-16

DAILY TASK 3 - GAIN CALIBRATION - STEP 1

• Allow up to 30 mins for 

Airo

 to reach its operating 

temperature.

• Select 

Hospital Service

 from the 

SERVICE 

FUNCTIONS

 screen.

• Enter password as prompted.
• Select 

Gain Calibration

 

from the 

HOSPITAL 

SERVICE

 screen.

• Select 

Next

 to proceed when prompted.

NOTE: 

Ensure the field of view for 

Airo

 is clear.

DAILY TASK 2 - EMERGENCY STOP TEST - 

STEP 2

• Remove the pendant and move to a safe area.
• Press and hold the 

Scan

 button and wait for the 

gantry to move and beep tone to start.

• While holding the 

Scan

 button down, press the 

E-

Stop

 button on the pendant, ensuring the beeping 

tone stops within 0.5 seconds and the gantry stops 
immediately.

• Click 

Next

 to continue. Contact 

Brainlab

 if test fails.

DAILY TASK 3 - GAIN CALIBRATION - STEP 2

• Remove the pendant and go to a safe area, ensuring 

the scanning area is also safe for X-ray exposure.

• Select and hold the 

Scan

 button.

• Select 

Next

 if the Gain Calibration is successful.

If not successful, evaluate the area and remove any 
objects that might interfere with the scan and repeat 
scan.

DAILY TASK 4 - QUALITY CHECK SCAN

• Select 

Quality Check Scans

 from the 

HOSPITAL 

SERVICE

 screen.

• Select 

Daily Helical QC

 

a

 and/or 

Daily Axial QC

 

s

 (depending on clinical use).

• Position the phantom on the table top and align the 

lasers as prompted (use Carbon X-tra only).

• Ensuring perfect alignment, click 

Next

 to continue.

• Remove the pendant and move to a safe area.
• Press and hold the 

Scan

 button until the yellow light 

on the pendant goes out.

• Select 

Next 

and repeat as prompted for the 

opposite scan direction if the scan is successful.

Scan

a

a

s

Scan

Содержание airo2

Страница 1: ...rom the Choose Mode screen on the pendant Alternatively select Transport on the pendant by pressing Menu a Transport LATCH AIRO Ensure the gantry is at the end of the rail away from the table column R...

Страница 2: ...to people or damage to objects NOTE Inclines must not exceed 5 and thresholds cannot exceed 9 5 mm 3 8 inch PARK IN THE OR Select Scan a from the Choose Mode screen Press the Down s button on the pen...

Страница 3: ...also safe for X ray exposure Select and hold the Scan button Select Next if the warm up is successful DAILY TASK 2 EMERGENCY STOP TEST STEP 1 Select E Stop Test from the HOSPITAL SERVICE screen Follo...

Страница 4: ...ck Next to continue Contact Brainlab if test fails DAILY TASK 3 GAIN CALIBRATION STEP 2 Remove the pendant and go to a safe area ensuring the scanning area is also safe for X ray exposure Select and h...

Страница 5: ...ation immediately prior to scanning Press and hold the Scan button during scanning and follow prompts from the pendant OPERATING POSITION 1 NO ROTATION Move gantry f to full extent of the base and awa...

Страница 6: ...Note This guide does not replace the user manuals Page 6 16...

Страница 7: ...tion immediately prior to scanning Press and hold the Scan button during scanning and follow prompts from the pendant OPERATING POSITION 1 ROTATED 180 DEGS Move gantry g to the end of the rail and awa...

Страница 8: ...head of the table Position the patient on the table with the head facing towards the camera Remove the pendant and go to a safe area Halt respiration immediately prior to scanning Press and hold the S...

Страница 9: ...away from the camera Remove the pendant and go to a safe area Press and hold the Scan button during scanning and follow prompts from the pendant OPERATING POSITION 1 ROTATED 90 DEGS Move gantry g to t...

Страница 10: ...Note This guide does not replace the user manuals Page 10 16...

Страница 11: ...ansfer pedal d on the shuttle or the Zero Position key on the remote control The column moves to the upper lift position NOTE Refer to the table top instruction manual for more information STEP 3 Tran...

Страница 12: ...Note This guide does not replace the user manuals Page 12 16...

Страница 13: ...ro Within Spine Trauma Navigation Select Register New Scan Airo Scanner NOTE No user interaction on Curve is required during the registration workflow STEP 3 Enter the patient s weight as prompted Air...

Страница 14: ...Navigation to perform a scan with without automatic registration If Navigation is not checked the data will not automatically be registered for navigation Select Next The system automatically verifies...

Страница 15: ...r X ray exposure Press and hold Scan s for the scan duration SET 3D SCAN AREA Define the 3D scan area d using Left Bound Right Bound and Field of View a Select Next f NOTE Yellow lines s indicate the...

Страница 16: ...ODULATION IF REQUIRED NOTE This is only applicable if helical scan mode is chosen and an A P scout has been performed NOTE The minimum maximum and average tube current values based on the previous A P...

Отзывы: