Note: This guide does not replace the user manuals.

Page 1-16

POWER ON AIRO

• Insert the key and switch on the system by rotating 

the key to the 

On

 position 

d

.

• Press  the 

Start

 

button.

The yellow indicator light 

illuminates.

 

REMOVE TABLE TOP

• Ensure the table top is removed.

TRANSPORT

• Select 

Transport

 from the 

Choose Mode

 screen 

on the pendant.

• Alternatively, select 

Transport

 on the pendant by 

pressing 

Menu

 

Transport

.

LATCH AIRO

• Ensure the gantry is at the end of the rail, away from 

the table column.

• Rotate the gantry and align it with the base. 

Ensure the latch is locked.

• Push the gantry towards the table column until it also 

latches with the base.

NOTE: 

Ensure that the gantry does not come into 

contact with the table column during alignment.

a

Transport

a s

d

...............................................................................................................................................................................................................................................

Hardware: Airo

MOVE, PARK AND STORE AIRO

Содержание airo2

Страница 1: ...rom the Choose Mode screen on the pendant Alternatively select Transport on the pendant by pressing Menu a Transport LATCH AIRO Ensure the gantry is at the end of the rail away from the table column R...

Страница 2: ...to people or damage to objects NOTE Inclines must not exceed 5 and thresholds cannot exceed 9 5 mm 3 8 inch PARK IN THE OR Select Scan a from the Choose Mode screen Press the Down s button on the pen...

Страница 3: ...also safe for X ray exposure Select and hold the Scan button Select Next if the warm up is successful DAILY TASK 2 EMERGENCY STOP TEST STEP 1 Select E Stop Test from the HOSPITAL SERVICE screen Follo...

Страница 4: ...ck Next to continue Contact Brainlab if test fails DAILY TASK 3 GAIN CALIBRATION STEP 2 Remove the pendant and go to a safe area ensuring the scanning area is also safe for X ray exposure Select and h...

Страница 5: ...ation immediately prior to scanning Press and hold the Scan button during scanning and follow prompts from the pendant OPERATING POSITION 1 NO ROTATION Move gantry f to full extent of the base and awa...

Страница 6: ...Note This guide does not replace the user manuals Page 6 16...

Страница 7: ...tion immediately prior to scanning Press and hold the Scan button during scanning and follow prompts from the pendant OPERATING POSITION 1 ROTATED 180 DEGS Move gantry g to the end of the rail and awa...

Страница 8: ...head of the table Position the patient on the table with the head facing towards the camera Remove the pendant and go to a safe area Halt respiration immediately prior to scanning Press and hold the S...

Страница 9: ...away from the camera Remove the pendant and go to a safe area Press and hold the Scan button during scanning and follow prompts from the pendant OPERATING POSITION 1 ROTATED 90 DEGS Move gantry g to t...

Страница 10: ...Note This guide does not replace the user manuals Page 10 16...

Страница 11: ...ansfer pedal d on the shuttle or the Zero Position key on the remote control The column moves to the upper lift position NOTE Refer to the table top instruction manual for more information STEP 3 Tran...

Страница 12: ...Note This guide does not replace the user manuals Page 12 16...

Страница 13: ...ro Within Spine Trauma Navigation Select Register New Scan Airo Scanner NOTE No user interaction on Curve is required during the registration workflow STEP 3 Enter the patient s weight as prompted Air...

Страница 14: ...Navigation to perform a scan with without automatic registration If Navigation is not checked the data will not automatically be registered for navigation Select Next The system automatically verifies...

Страница 15: ...r X ray exposure Press and hold Scan s for the scan duration SET 3D SCAN AREA Define the 3D scan area d using Left Bound Right Bound and Field of View a Select Next f NOTE Yellow lines s indicate the...

Страница 16: ...ODULATION IF REQUIRED NOTE This is only applicable if helical scan mode is chosen and an A P scout has been performed NOTE The minimum maximum and average tube current values based on the previous A P...