FGC100 User Manual 

 

12 

 

 

5.3

 

Calibration of the spectrometer only 

  

Follow these steps: 

 

To calibrate the FluxGage click ‘Calibration’ on the top banner of the FluxGage software. 

 

In the setting area, enter the total luminous flux value zero 

 

Click ‘Open Light File’ and select the spectral calibration file provided with the Disk-On-
Key included in the FGC100. 

 

If the file is loaded correctly, the ‘calibration file valid’ indicator will turn green. 

 

Press the push button to turn the FGC100 lamp unit on. 

 

Press ‘Autoset’ and wait for the operation to complete. 

 

In the setting area, select 30 averages.  

 

Press the push button again to turn the FGC100 lamp unit off. 

 

Press ‘Background’ and wait for the operation to complete. 

 

Press the push button to turn the FGC100 lamp unit on, and allow enough time to reach 
thermal stabilization (~30minutes). 

 

Press ‘Calibrate’ and wait for the operation to complete. 

 

At the top of the screen, the measured flux of the calibration source will appear. It 
should match the value entered in the settings area. 

 

Click on the top right corner to exit the window. 

 

 

5.4

 

Back up calibration 

 

In addition to the working calibration there is a backup calibration. 

The backup calibration can be retrieved by pressing ‘restore back up calibration’. This will copy 
the backup calibration into the working calibration. The backup calibration can be overwritten 
by pressing ‘over write back up calibration’. This should be done when the calibration is done 
with an Ophir certified calibration unit. 

 

 

 

 

 

 

 

 

 

FGC100 User Manual

 

 05 February 2019

 

Rev 1.01-2

 

 

Содержание Ophir FluxGage FGC100

Страница 1: ...Ophir Optronics Solutions Ltd FGC100 Calibration Unit FGC100 User Manual FluxGage FluxGage...

Страница 2: ...FGC100 7 4 1 Operating the FGC100 9 5 Performing Calibration of FluxGage Using FGC100 10 5 1 Calibration of the spectrum measurement and the total flux measurement 11 5 2 Calibration of the total flux...

Страница 3: ...complies with standard IEC 62471 2006 First Edition Photobiological safety of lamps and lamp systems and is classified as Low risk Risk Group 1 when used as a standalone luminaire When the FGC100 cal...

Страница 4: ...D type cable 100 230VAC mains cable Disk on key Lamp Unit Control Unit 12VDC Adapter 100 230VAC Mains D type Cable Figure 1 FGC100 Block Diagram The lamp unit includes LED mounted on a heat sink Tempe...

Страница 5: ...ature stabilization circuit General on off switch Hour meter LED on off push button LED temperature indicator light Green Red Back Panel D type connector USB connector for usage at Ophir service cente...

Страница 6: ...FGC100 User Manual 6 3 Mechanical Drawings...

Страница 7: ...e the round cover from the lid Remove the protective cover from the FGC100 lamp unit Insert the FGC100 lamp unit into the hole in the lid and secure it using the two thumb screws Connect the D type ca...

Страница 8: ...FGC100 User Manual 8 Figure 5 FGC100 connected to FuxGage...

Страница 9: ...lamp unit The temperature indicator will be red initially and will turn to green after a few minutes when the temperature of the lamp unit will be stabilized However wait at least 30 minutes before u...

Страница 10: ...FGC100 User Manual 10 5 Performing Calibration of FluxGage Using FGC100...

Страница 11: ...the top of the screen the measured flux of the calibration source will appear It should match the value entered in the settings area Click on the top right corner to exit the window 5 2 Calibration o...

Страница 12: ...nd wait for the operation to complete Press the push button to turn the FGC100 lamp unit on and allow enough time to reach thermal stabilization 30minutes Press Calibrate and wait for the operation to...

Отзывы: