CONEX-LDS
Controller GUI Manual
Start performance verification Button
Opens a user-friendly interface to validate the CONEX-LDS calibration. The label on
the button changes to “Stop” to cancel the current calibration verification process.
•
Error
≤
3%
=
GREEN
indicator, OK
•
Error >
3%
=
RED
indicator, consider sending the CONEX-LDS to the factory for
calibration
REFERENCE INFORMATION
The optical wedge of the CONEX-LDS-VER calibration verification kit induces
an angular deviation to the autocollimator's beam. This deviation corresponds to
a measurement angle of approximately 1000 µrad. The exact angle is measured
and reported in the Calibration Certificate shipped with each CONEX-LDS-
VER Kit. The user should input KitSerialNumber and ReferenceAngleValue
from the Calibration Certificate in the Configuration tab. The verification
process consists of 4 measurements, spaced 90° apart on the PSD. These
measurements are compared to the CONEX-LDS-VER reference angle set by the
user in the Configuration tab. Before performing this verification and inserting
the optical wedge, the mirror must be centered so the CONEX-LDS reads X = 0
and Y = 0 ±10 µrad.
11
EDH0331En1032 — 12/18