MKS MicroPirani 925 Series Скачать руководство пользователя страница 3

Use a cable with a female 6-pin connector with strain reliefs to ensure proper 
electrical connection and to reduce stress.  

 

 

Ensure a low impedance electrical connection between the 
901P transducer body and the grounded vacuum system to 
shield the sensor from external electromagnetic sources.  

 

To comply with EN61326-1 immunity requirements, use a braided, shielded 
cable. Connect the braid to the metal hoods at both ends of the cable with 
the end for power supply connected to earth ground. 
 

925 I/O Connector (6 pin) 
 
 

 

 

6 pin Hirschmann 

 

 
 

PIN 

Description 

1  Identification resistor 3KΩ 
2  Analog  
3  Analog Output – 
4  Power + (9-30 VDC) 
5  Power return - 
6  Chassis 
 

 

 
 

 

 
 
  
 