P/N: 100017410
Short form manual
REV: A, December 2009
Product: 925-x1090-0051
Connector: 6 pin Hirschmann
Output: 19 (TPR280,TPR265)
RoHS Compliant
Measuring range (N
2
and Air):
1×10
-5
Torr to Atmosphere
Accuracy
(1)
(N
2
)
5×10
-4
to 1x10
-3
Torr:
±10% of reading
1×10
-3
to 100 Torr:
± 5% of reading
100 to Atm.:
± 25% of reading
Repeatability
(1)
(N
2
): 1×10
-3
to 100 Torr
± 2% of reading
Supply Voltage:
9 – 30 VDC
Power consumption:
< 1.2 Watt
Fuse (thermal recoverable):
200 mA
Analog output 1: (100 Ω impedance):
2.199 – 8.625 VDC
Analog output 1 resolution:
16 bit
Materials exposed to vacuum:
304 Stainless steel, Silicon
SiO
2
, Si
3
N
4
, gold, Viton
® (2)
Low out gassing Epoxy resin
Internal volume:
KF16 flange
2.80 cm
3
KF16 long flange
6.63 cm
3
KF25 flange
3.55 cm
3
VCR4 flange
1.84 cm
3
VCR8 flange
3.06 cm
3
Housing material:
Stainless steel 304
Weight (925 with KF16):
170g
Operating temperature:
0 to 40
o
C
Bakeout temperature (Power off):
85
o
C
(1) Accuracy and repeatability values are typical values measure in Nitrogen atmosphere at
ambient temperature after zero adjustment.
(2) Transducer without Viton seal is available for use in UHV applications.
MKS Instruments HPS Products
www.mksinst.com