![MKS HPS 919 Series Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/mks/hps-919-series/hps-919-series_operation-and-maintenance-manual_1813430027.webp)
20
Hot Cathode Ionization High Vacuum Senor System
Sensor
Type
Bayard-Alpert Low Power Nude
Sensitivity
9 Torr
-1
(±20%)
Degas Power
40 W max
Operating Voltage
Grid
180 VDC
Filament bias
30 VDC
Filament
4.7 VAC @ 1.8 A
Calibration Gas
Air/nitrogen
Installation Orientation
Any
Materials Exposed to Vacuum
304 SS, Inconel
®
X-750, glass,
tungsten, platinum, nickel, either
yttria-coated iridium or tungsten
Operating Temperature Range
0° to 50°C (32° to 122°F)
Bakeout Temperature
300°C max, with CF
150°C max, with KF and Viton
®
Typical Weight (with CF Flange)
0.9 lb (0.4 kg)
Vacuum Connection
2¾" CF (non-rotatable)
KF 40
Design and/or specifications subject to change without notice.