Chapter Two: Installation
Setup
51
How to Wire a PDR-C and PDR-D Series Readout to a Type 722C Transducer with
a 9-
pin Type “D” Connector.
Caution
DO NOT use the -15 VDC output of the PDR-C and PDR-D readout
to power the 722C transducer. When the -15 VDC signal of the PDR-
C and PDR-D is connected to the power return of the transducer, a
short between the -15 VDC of the transducer and the A GND of the
PDR-C and PDR-D occurs. The PDR
–
C and PDR-D readout will
blank out. Either unit may be damaged.
The correct cable to use to connect the 9 pin Type “D” connector to the PDR
-C and PDR- D
readout is a RCB700-
2 cable. This cable has the mating Type “D” connector on the transducer
end and flying leads on the PDR-C and PDR-D end. Table 4 lists the pin assignments for the
various colored wires on the flying leads end of the cable.
Connections Between a Transducer with a Type “D” Connector and a PDR
-C/-D Readout
Pin Number
Transducer Signal
Flying Lead Color Code
PDR Signal
1
Pressure Output
Red
Pressure Input
4
+ Power Input
Green
+ 15 VDC
8
Pressure Return
Black
Signal Ground (A GND)
9
Power Return
White
Power Ground (D GND)
Table
8: Connections Between a Transducer with a Type “D” Connector and a PDR
-C /- D Readout
Содержание Baratron 722C
Страница 1: ...20054397 001 Rev A 11 21 MKS Baratron Type 722C Absolute Pressure Transducer...
Страница 5: ...Table of Contents v Index 71...
Страница 6: ......
Страница 8: ...List of Figures and Tables viii...
Страница 58: ...How To Check the Transducer Zero Chapter Two Installation 58 This page intentionally left blank...
Страница 62: ...General Chapter Four Operation 62 This page intentionally left blank...
Страница 70: ...NW 16 KF Fitting Information Appendix B Model Code Explanation 70 This page intentionally left blank...