Chapter Two: Installation
Setup
17
Attaching the Terminal Block Connector Cable
The cable to the terminal block connector must be firmly attached to the top of the transducer, in
order to comply with CE Mark requirements. Figure 8 shows the preferred method to connect
the cable; using a metal cable clamp to affix the cable to the transducer. Figure 9 shows an
alternate method; wrapping the braided shield under the screw. Both examples show the cable
securely clamped to the transducer.
Transducer
Overall Insulation
(if present)
Bare Metal Cable Clamp
Making Firm Contact To Braid
Braid Here Is Desirable
(but not usually necessary)
Keep Wires and Braid
Flat Against Case
Metal Cable Clamp
Screw
Split Lock Washer
External Tooth Lock Washer
Transducer Housing
Optional Plastic or Metal Cable
Clamp (For Physical Strain Relief)
Figure 8: How To Connect the Cable to a Terminal Block (Example 1)
Transducer
Overall Insulation
(if present)
Braid Strands Twisted
Under Washer
Keep Wires Flat
Against Case
Optional Plastic or Metal Cable
Clamp (For Physical Strain Relief)
Screw
Split Lock Washer
Flat Washers
External Tooth Lock Washer
Transducer Housing
Figure 9: How To Connect the Cable to a Terminal Block (Example 2)