Chapter Two: Installation
Setup
9
Setup
Mounting
The 722 transducer can be mounted with the cylindrical end in either a vertical (upright) or
horizontal position. The mounting requirements allow any foreign matter entering the pressure
port to fall
away from
the sensing diaphragm.
Mounting the Unit in a Vertical Position
If the unit is mounted in a vertical position, the cylindrical end of the unit must point upwards, as
shown in Figure 5. Do not install the unit with the cylindrical end of the unit pointing downward
because particulate impurities may accumulate on the sensing diaphragm and alter the pressure
reading.
4-VCR Female
shown
4-VCR Female
shown
Acceptable Vertical Mounting
Unacceptable
Vertical Mounting
Figure 5: Vertical Mounting Positions
Mounting the Unit in a Horizontal Position
In a horizontal position, the cylindrical end of the unit can point in any direction.
4-VCR Female
shown
Figure 6: Acceptable Horizontal Mounting Position