Appendix A: Product Specifications
Electrical Specifications
35
Appendix A: Product Specifications
Electrical Specifications
CE Compliance
Electromagnetic Compatibility
1
EMC Directive 89/336/EEC
Input Power Required
±15 VDC @ 35 mA, regulated ±5%
Output Signal
0 to 10 VDC into 10K ohm or greater
Performance Specifications
Accuracy
2
± 0.5% of Reading ± temperature coefficients
(optional ± 0.15% available on 10, 100, and 1K units
only)
Full Scale Ranges (Torr)
1, 2, 10, 100, 1000, 5000, 10000, 15000, 20000, 25000
Materials Exposed to Gases
Inconel and stainless steel
Overpressure Limit
120% of FS or 35 psia (240 kPa), whichever is greater
Resolution
0.01% FS
Temperature Coefficients
Zero
Span
± 0.008 % of FS/° C
± 0.04% of Reading/° C
Volume (Px side)
1, 2, 10, 100, 1000 Torr
5k, 10k, 15k, 20k, 25k Torr
7.00 cc
8.5 cc
Physical Specifications
Fittings
Standard
Optional
½” (12.7 mm) tubulation < 1000T
Swagelok
®
8-VCR
®
only on 5K and higher ranges
KF-16, Swagelok 8-VCR, Swagelok 8-VCO
®
1
An overall metal braided shielded cable, properly grounded at both ends, is required during use.
2
Includes non-linearity, hysteresis, and non-repeatability.
Содержание Baratron 122B
Страница 4: ...110899 P1 Rev D 8 98 MKS Baratron Type 122B Absolute Pressure Transducer ...
Страница 9: ...List of Figures and Tables vi ...
Страница 37: ...Sensor Chapter Three Overview 28 Figure 7 Cross Sectional View of the Type 122 Transducer ...
Страница 43: ...Sensor Chapter Five Maintenance 34 This page intentionally left blank ...