Error codes
Symptom / mA signal:
Fault:
Remedy:
Constant current signal
Electrical connection Check connection and make the necessary
or I = 0 mA
is not made correctly. corrections.
Unsteady current signal
Undervoltage.
Check the voltage at transmitter.
Min. voltage @ max. load > 12 V DC!
I = 3.80 mA
Undervoltage.
Check the voltage at transmitter.
Min. voltage @ max. load > 12 V DC!
I = 3.85 mA
Electrode error.
Exchange the electrode.
I = 3.90 mA
Calibration
not possible.
Exchange the electrode.
I = 3.95 mA
pH lower than
Readjust the measuring range to a lower range.
measuring range.
I = 4.00 mA
Error on startup.
Disconnect the transmitter from PSU for
min. 5 seconds and try again.
I = 22.00 mA
pH higher than
Readjust the measuring range to a higher range.
measuring range.
Standard transmitter configuration
The transmitter is standard configured as described below on delivery:
pH transmitter, item no. 203111:
Measuring range: 0-14 pH = 4-20 mA, 0-point = 4,6 pH
pH and temperature transmitter, item no. 203112:
Measuring range: 0-14 pH = 4-20 mA, 0-punkt = 4,6 pH, Temperature: 0-50° C = 4-20 mA
Redox transmitter, item no. 203121:
Measuring range: -1000 to +1000 mV = 4-20 mA
HART
®
commands:
All commands and settings can be read and set by means of HART
®
commands. Ask for
separate documentation regarding HART
®
commands.