13
Mounting
SW 836502
Electrical mounting
General
pHix
®
Compact
must not be installed in
explosion hazardous locations!
pHix
®
Compact
should be connected to an
active 4-20 mA input or be supplied from a
separate 12 - 30 VDC voltage source.
Max. power consumption is 50 mW.
pHix
®
Compact's
4-20 mA output(s) is gal-
vanically separated from the liquid.
Signal cable
pHix
®
Compact
comes with either 1 or 2
galvanically separated signal outputs.
pHix
®
Compact
with 1 x 4-20 mA output for
ph or redox measurement.
pHix
®
Compact
with 2 x 4-20 mA outputs, 1 for
ph or redox measurement and 1 for temperature
measurement.
➀
➁
➂
➀
➁
➂
1
1
2
2
➀
➁
➂
➃
➄
1
1
2
2
PVC Cable
PVC Cable
PUR Cable
-> 2005
2006 -> 07
2007 ->
Nr.:
Designation:
Color:
Color:
Color:
Signal:
1
Shield
2
+ 12-30 V DC
Red
Red
1 Black
pH- or redox signal
3
- 4-20 mA
Black
Black
2 Black
Output 1
PVC Cable
PVC Cable
PUR Cable
-> 2005
2006 -> 07
2007 ->
Nr.:
Designation:
Color:
Color:
Color:
Signal:
1
Shield
2
+ 12-30 V DC
Red
Red
1 Black
pH- or redox signal
3
- 4-20 mA
Black
Black
2 Black
Output 1
4
+ 12-30 V DC
Yellow
Orange
Grey
Temperature signal
5
- 4-20 mA
Orange
Brown
Brown
Output 2
Note:
PHix Compact is delivered with
cables made of PVC or PUR.
Check the type of cable according
to the specifications before con-
necting pHix Compact.