9
2
Before use
- Do not pull the control unit cord
forcefully.
- Do not apply strong shock to the
control unit.
It may get damaged.
CAUTION
- Make sure to attach the provided mattress securely. Improper attachment may cause a patient to get injured.
- Position a patient's body at least 10 mm (0.39 in) away from the side rails. When using the electric
scalpel, they may become hot and if the body touches the rails, it may cause the patient to get burned.
Positioning the patient
- Follow the steps below to position the patient.
1. Attach the accessory mattress onto the tabletop with their Velcro fastenings aligned with each other.
2. Put the patient on the mattresses.
3. Position the patient according to the purpose of the surgical operation.
Prohibition of disassembling and modification
- Do not attempt to disassemble or modify the table. It
may cause a malfunction.
Patient’s body position
- When using accessories to secure a patient's body position,
make sure to monitor the condition of the patient. Improper
body position can lead to paralysis or pressure injuries.
When uses together with other medical electrical equipment
- When using a high-frequency surgical equipment and/or
a cardiac defibrillator etc. with the operating table, refer
to their operator's manual provided by the manufacturers
and observe the precautions and usage. Improper
precautions and usage may cause the operator or the
patient to get burned and/or devices to malfunction.
- Before using together with other equipment, make
sure to check that it is not affected by electromagnetic
interference.
An electromagnetic interference from other medical
electrical equipment may cause malfunction.
Allowable patient body weight
- Do not apply loads which exceed the allowable patient
body weight (total of the patient and accessories)*. The
table may become inoperable and cause malfunction.
*Lift: 450kg (992lbs)/ Trendelenburg, Lateral tilt: 360kg (794lbs)
Preventive maintenance and inspection
- Make sure to perform inspection before and after use,
and appropriate preventive maintenance inspection.
The operating table may require replacement of
the parts due to significant wear, deterioration, and/
or breakage depending on the length of service and
frequency of use.
- For preventive maintenance inspection, please contact
your distributor or us directly.
Antistatic measure
- Do not use the operating table on floors and/or with
accessories that do not possess static electricity
countermeasures. This may impede surgical operations.
WARNING
uneven floor surface
pad
Installation of the operating table
Patient's position during surgical operation
Others
- Have the person who operates
the operating table at the position
where the emergency stop switch
can be immediately pressed, and
the patient’s condition can be
constantly observed.
CAUTION
Operator's position
Back plate
Head plate
Operator's position
180°
19-
11/
16”
(50
0 m
m)
Head of the patient
WARNING
Control unit
Содержание MOT-VS600
Страница 2: ......
Страница 4: ......
Страница 5: ...1 Introduction 1 Introduction ...
Страница 6: ...2 1 Introduction ...
Страница 10: ...6 1 Introduction ...
Страница 11: ...2 Before use 2 Before use ...
Страница 19: ...3 Component identification 3 Component identification ...
Страница 24: ...20 3 Component identification ...
Страница 60: ...56 4 Operation ...
Страница 61: ...5 Maintenance and inspection 5 Maintenance and inspection ...
Страница 64: ...60 5 Maintenance and inspection ...
Страница 65: ...6 Specifications 6 Specifications ...
Страница 69: ...7 Troubleshooting 7 Troubleshooting ...
Страница 71: ...8 Before contacting for repairs 8 Before contacting for repairs ...