Flexible Magnetic Stand No. 7012-10
Manual No.99MBH007B1
------------------------------------------------------------------------------------
If you use a pacemaker, you must not use the Magnetic Stand.
The magnetic force on the Magnetic Stand may cause
malfunction of your pacemaker.
------------------------------------------------------------------------------------
------------------------------------------------------------------------------------
・
Do not bring the Magnetic Stand near the floppy disk or other
data storage media. Otherwise the data stored in these media
may be deleted due to the magnetic force.
・
The attractive force of a magnetic stand will be reduced
depending on the conditions of a surface the magnetic base
attracts, including material, thickness, surface roughness, and
painting applied.
・
The attractive force also reduces if the magnetic base
attraction surface has scratches or rust on it.
・
Dropping or bumping the magnetic stand will cause a
breakage or defective in the instrument.
・
When fixing the Magnetic Stand on the thick-coated surface,
make sure that the instrument is securely fixed on the surface
to avoid breakage or personal injury by dropping of the
Magnetic Stand.
・
To prevent deterioration in holding power of magnet, avoid
using the fixture in the proximity of transformers or
demagnetizer as they generate strong magnetic field, affecting
the fixture
’
s performance. Be sure to keep the magnet base
lever OFF when not in use.
・
Do not disassemble the arm and the magnet base, or its
performance can be deteriorated.
------------------------------------------------------------------------------------
1. Features
This magnetic stand is a measuring fixture for setting up a light-weight, low stylus
force measuring tool such as a lever-type dial indicator (test indicator) in an
appropriate measuring position.
・
Employs the magnet base permitting the fixture to be mounted either horizontally,
vertically, or even upside down. Also, a V-groove on the bottom of the base
permits the fixture to be mounted even on something cylindrical.
・
The pole can be locked at any desired position with the tension of the wire.
2. Name of each part (see Fig. 1)
①
Magnet base
②
Lever
③
Handle
④
Wire adjuster
⑤
Pole
⑥
Adjusting sleeve
⑦
Lock nut
⑧
Mounting holder
⑨
Tightening nut
⑩
Bushing
⑪
Retainer ring
Fig. 1
3. Usage
Please grasp the clamping function according to the handle position prior to use
(see Fig. 2).
1) Ensure that the handle is in free position and that the adjusting sleeve and the
locknut are not in close contact with each other.
2) To clamp the pole, rotate the adjusting sleeve to temporarily adjust it so that the
pole can be clamped with the handle at the clamping position. (see Fig. 3)
3) Insert the stem of a measuring tool in the mounting holder hole, then
temporarily clamp it with the tightening nut.
4) Locate the magnetic base in place so that the tip (contact point) of the
measuring tool comes into contact with the measured point, then turn the lever
ON to have the magnetic base fixed by attraction.
5) Adjust the degree of pole flexure and the measuring tool orientation so that the
pole can be clamped with the handle at the clamp position while rotating the
adjusting sleeve. (see Fig.4)
6) Rotate the locknut to bring it into close contact with the adjusting sleeve in
order to prevent the sleeve from loosening/
7) Ensure that the measuring tool has come in contact with the measured point in
an appropriate orientation, clamp the measuring tool with the tightening nut,
and then clamp the pole by turning the handle to the clamp position.
IMPORTANT
--------------------------------------------------------------------------------------------------------------
・
A measuring tool with a stem of less than 16 mm in length may not be fixed firmly
since this tool can be in an inclined orientation.
・
To minimize the possible measuring errors, fix the measuring tool in an appropriate
orientation, referring to the Users Manual of the measuring tool.
・
Hold the measuring tool mounted on the magnetic stand by hand when loosening
the handle.
-----------------------------------------------------------------------------------------------
Fig. 2
Fig. 3 Fig. 4
------------------------------------------------------------------------------
・
Do not perform undue adjustment with the pole being
clamped. This will cause a breakage in the clamp wire.
・
Do not apply the excessive force to the handle when you
clamp the pillar.
------------------------------------------------------------------------------
4. Maintenance part
Part No.
Part name
Remarks
02AZC410 Post
assemble One set of
⑤
(including
⑥
and
⑦
)
of Fig .1
●
Use the procedures below for replacing the broken wire:
1) Remove parts
⑧
,
⑨
, and
⑩
shown in Fig.1.
2) Remove
⑪
using a pair of pliers or the like and take out
③
.
3) Remove a set of parts
④
and
⑤
, install the replacement assembly to
④
,
then reverse the procedures to restore the assembled state.
5. Specifications
Code No.
Applicable stem dia.
Mass
Magnetic force
7012-10
φ
6,
φ
8,
φ
9.53
(
3/8"
)
1.4kg 600N
WARNING
CAUTION
CAUTION