32
MIR Spirolab (UK version) | User Manual
Troubleshooting
Problem
Message
Possible Cause
Solution
Spirolab doesn’t turn on
No Message
Battery pack may be flat
Charge the device using the charger
unit
No Message
The device may have lost
internal software
Connect the device to a PC via USB
and update the software. For further
information contact Intermedical
service department.
Problems turning the device on
‘Ram error
data recovery
Please wait’
The data in the device memory
has been damaged
If the data has been correctly
restored the last standard turning on
process will be repeated. If not
Contact Intermedical service depart-
ment.
The device turns itself off then
reboots while being used.
No message
There is an internal error
Connect the device to a PC via USB
and update the software. For further
information contact Intermedical
service department.
On completion of spirometry
tests, the readings are not
credible
No message
The turbine may be dirty
Clean the turbine as described in
section 4.1; if necessary, replace the
turbine with a new one
No message
The test was performed in the
wrong manner
Repeat the test
On completion of a spirometer
test, some parameters are not
displayed
No message
Parameters not selected in cus-
tomise menu.
Large number of parameters
selected so cannot all be viewed
on screen
Check ‘Customise’, ‘Spirometer’,
‘Indices’ for selected parameters and
amend accordingly
Scroll down on the results to view all
parameters
During an oximeter test the
readings displayed are irregular,
intermittent or wrong
No message
The sensor is wrongly
positioned or the patient’s
perfusion is poor
Reposition the oximeter sensor
No message
The patient moved
For accurate results, the patient must
not make any sudden movements.
The display is not very bright
during tests
No message
The brightness level of the display
automatically dims 5
minutes after a test starts. This
function extends battery life.
None
Содержание Spirolab
Страница 1: ...spirolab UKVersion User Manual...
Страница 5: ......