The first few seconds of a test are used for finding the strongest signal. Once this has been
found, the timer resets itself and the Spirolab starts recording data. If the sensor is not
inserted correctly, icon transforms into and the icon appears alongside it. At the
same time the Spirolab will beep (if set to do so in the customise menu).
If the sensor has been inserted but the finger is not positioned correctly, the icon
transforms into and the icon appears alongside it. At the same time the Spirolab
will beep (if set to do so in the customise menu). By tapping the icon you can pause
the alarm for a few minutes.
If the alert situation persists then the beep will return after a few minutes
If the signal is received properly by the sensor, after a few seconds the device will begin to
beep and display readings on the screen.
During the test if the %SpO2 or HR rate goes above or below the threshold as set in the
customise menu then the Spirolab emits a beep (if set to do so in the customise menu) for
as long as this anomaly continues.
When the test ends then the results are displayed and the minimum, maximum and
average %SPO2 and HR is shown. The results can be saved and printed.
20
Perform Test
MIR Spirolab (UK version) | User Manual