background image

 

Spirobank II

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Rev.1.6 

Page 6 of 35 

User manual 

1.2.2

 

Turbine 

 

Disposable turbine

 

 WARNING 

For spirometry testing with  a disposable turbine it is important to use a new 

turbine  for  each  new  patient.  The  accuracy  and  hygiene  of  the  disposable 
turbine  can  only  be  guaranteed  if  it  has  been  conserved  beforehand  in  its 
original sealed packaging 
The  disposable  turbine  is  made  of  plastic  and  its  disposal  after  use  should 
adhere to the local regulations and norms in force. 

 

Reusable turbine 

 WARNING 

The correct functioning of the re-usable turbine can only be guaranteed if it has 
been cleaned in the correct manner and is free from foreign bodies which could 

alter its movement. If the turbine has not been cleaned sufficiently this could 
cause  cross-contamination  from  one  patient  to  another.  Periodic  cleaning 
should only be done when the device is for personal use and will only be used 
by one patient. The cleaning of the turbine should be performed according to 

the instructions contained in the User’s Manual.

 

 
The following information applies to both types of turbine. 
The turbine must never be held under running water or direct air pressure and must never come into contact with hot fluids.  
Do not allow dust or foreign matter to enter the turbine sensor which may alter the correct functioning and possibly cause damage. 

The presence of any impurities such as hair, sputum, threads etc. within the body of the turbine sensor may seriously compromise 
measurement accuracy. 
 

1.2.3

 

Mouthpiece 

 

Any  disposable  mouthpieces  included  with  the  spirometer  are  only  to  be  used  as  a  reference  guide  to  purchase  the  correct  size 
mouthpiece  required.  These  mouthpieces  are  clean  but  not  sterile.  To  purchase  appropriate  mouthpieces,  generally  either paper  or 
plastic, single-use/disposable, we suggest that you contact your local distributor. 

 

 

WARNING  

Use  a  bio-compatible  mouthpiece  to  avoid  any  problems  to  the  patient;  unsuitable  materials  could  cause  the  device  to 
malfunction, consequently providing incorrect test results. 

 

The user is responsible for obtaining the proper mouthpieces for the device. The required mouthpiece is a standard type with an outside 
diameter of 30 mm, is of common use and in general easily procured.  
 

 

WARNING 

To avoid environmental contamination caused by the disposal of used mouthpieces, the user must follow all the relevant local 
regulations. 

 

1.2.4

 

Oximetry sensors 

 

In addition to the sensor code 919024_INV supplied with the device, the following oximetry sensors can be used with 

Spirobank II: 

 

Manufacturer 

Code 

Description 

MIR code 

Envitec 

RS-3222-12 

Reusable small soft sensor (paediatric)  

939006 

Envitec 

RM-3222-12 

Reusable medium soft sensor (adults ) 

939007 

Envitec 

R-3222-12 

Reusable large soft sensor (adults) 

939008 

BCI 

3044 

Reusable hard finger sensor (adults)  

919020 

 

These sensors, with the exception of the sensor MIR code 919020 which has the MIR connector with orange arrow, require the use of 

an extension cable for a proper connection to 

Spirobank II

. Two cable lengths are available: 

 

 

Cod. 919200_INV 

 

length 1.5 m 

 

Cod. 919210_INV 

 

length 0.5 m 

 

Prolonged  use  and/or  the  patient’s  condition  may  require  changing  the  sensor  site  periodically.  Change  sensor  site  and  check 

skin 

integrity, blood circulation, and correct sensor alignment at least every 4 hours.  
 

 
 

Содержание Spirobank II

Страница 1: ...Spirobank II Rev 1 6 Page 1 of 35 User manual Spirobank II User manual rev 1 6 Issue date 20 01 2021 Approval date 20 01 2021...

Страница 2: ...HE Spirobank II 17 2 1 Switch on and switch off the device 17 2 2 Energy saving 18 2 3 Main screen 18 2 4 Symbols and Icons 18 2 5 Service menu 19 2 5 1 Reusable turbine calibration 22 2 6 Patient Dat...

Страница 3: ...al packaging In the unlikely event that you have a problem with your device please use the original packaging and return it to the distributor or manufacturer Should this be the case please follow the...

Страница 4: ...he patient by the doctor WARNING The manufacturer cannot be held responsible for any damage caused by the user of the device failing to follow instructions and warnings in this manual If the user of t...

Страница 5: ...Spirobank II is continuously checked during manufacturing and therefore the product complies with the established security levels and quality standards laid down by the Council Directive 93 42 EEC for...

Страница 6: ...spirometer are only to be used as a reference guide to purchase the correct size mouthpiece required These mouthpieces are clean but not sterile To purchase appropriate mouthpieces generally either pa...

Страница 7: ...vent safety hazards for the user 1 2 5 USB connection cable Incorrect use or application of the USB cable may produce inaccurate measurements wich will show very inaccurate values of the patient s con...

Страница 8: ...may be damaged or suffer a drop in overall performance The internal battery pack safety sensor could also be damaged as well by any of the above events Furthermore the user of the device could be har...

Страница 9: ...device or the battery charger and do not use it any longer as any of these events may cause acid leakage overheating smoke breakage and or fire NOTE The battery pack includes an internal safety protec...

Страница 10: ...ed against vertically falling drops of water Antenna symbol for devices that include RF transmitters FCC ID FCC Identification code indicating traceability to FCC compliance Rx ONLY Reference to US FD...

Страница 11: ...ielectric breakdown ESD can be caused by a buildup of static electricity by tribocharging or by electrostatic induction At lower relative humidity as the environment is drier charge generation will in...

Страница 12: ...nt to the next to ensure the maximum level of hygiene and safety For a correct interpretation of a spirometry test the measured values must be compared either to the so called normal or predicted valu...

Страница 13: ...y Flow at 75 of FVC L s FEV05 Volume expired after 0 5 seconds L FEV05 FEV05 FVC x 100 FEV075 Volume expired after 0 75 seconds L FEV075 FEV075 FVC x 100 FEV2 Volume expired in the first 2 seconds of...

Страница 14: ...usable soft sensor for adults Reusable paediatric soft sensor Range SpO2 Arms Range SpO2 Arms Range SpO2 Arms 70 100 1 19 70 100 1 470 70 100 1 390 70 80 0 554 70 80 1 626 70 80 1 851 80 90 1 32 80 90...

Страница 15: ...octor 1 6 3 Oximetry alarms description Spirobank II is equipped with audio and visual alarm indicators to alert the operator to provide prompt patient attention or to abnormal device conditions Spiro...

Страница 16: ...h of the above mentioned oximetry sensors is used 1 6 4 Other features Memory Memory capacity for over 10000 spirometric tests The precise number depends on the individual configuration so it cannot b...

Страница 17: ...discharge protection 8kV contact 15kV air Magnetic field immunity 30 A m Radio Frequency Immunity 3V m 80 2700 MHz MIR will make available on request circuit diagrams component part lists description...

Страница 18: ...displayed with a reduction of the indicators 2 3 Main screen On the main screen the following areas can be accessed patient data management area spirometry area oximetry area archive area archive are...

Страница 19: ...e key and then the key The service menu shows the following list of items Change date time LCD settings Bluetooth settings Select language Delete memory Select standard Select predicted Select turbine...

Страница 20: ...standard to be used ATS ERS or NHANES III with the keys and then press the setting takes effect and the device returns to the Service Menu WARNING If the NHANES III standard is selected it is not pos...

Страница 21: ...hreshold or rise above the maximum threshold of the SpO2 and BPM values previously set Use keys and to decrease increase the values and select moving the arrow with the keys and At the end press to re...

Страница 22: ...g of the turbine is necessary The disposable turbine is check before the application of the external packaging for this reason it doesn t requires a periodic calibration If a calibration must be perfo...

Страница 23: ...robank II is calibrated correctly then the FVC syringe value will be 3 00 FVC x 1 026 BTPS 3 08 L FVC at BTPS If the ambient temperature is 20 C the FIVC syringe value will be 3 00 FIVC x 1 102 BTPS 3...

Страница 24: ...en without modifying any data WARNING A new patient is not created from the previous patient when selecting this function Patient info however can be modified Future tests will be associated to the pa...

Страница 25: ...efault settings of the PC software will be transferred to the device and will remain in the device even when used in stand alone mode for example when the Spirobank II is connected to the PC and the u...

Страница 26: ...results of the best test To end the test press During the test the Spirobank II emits beeps the frequency of which are directly proportional to the inspired and expired velocity of the air This helps...

Страница 27: ...rst screen displays a Flow Volume graph of the Forced Vital Capacity pressing it is be possible visualize the best value of FVC FEV1 FEV1 and PEF with the percentage change compared to the predicted v...

Страница 28: ...argest FVC is 100 mL 2 11 Oximetry Testing WARNING Check if the oximetry function is available in the device this function is an option in some models This is not available in BASIC version WARNING Th...

Страница 29: ...the Spirobank II timer resets itself and the device starts recording data For any type of oximetry test if the sensor is not properly connected the following message will be displayed on screen after...

Страница 30: ...ct to biocompatibility tests Choose an application site on the patient s finger or toe where the light source will be directly over and in line with the detector The preferred sites are the forefinger...

Страница 31: ...manage Spirobank II The user can perform on line wireless tests of spirometry or oximetry and the tests will be directly stored in the database of winspiroPRO software 3 2 PC connection via USB port W...

Страница 32: ...liquid solution and shake it so as to remove any impurities Leave the turbine immersed for the time specified in the instruction of the solution WARNING To avoid irreparable damage to the reusable tur...

Страница 33: ...an the sensor WARNING Do not sterilize by irradiation steam or by using ethylene oxide Unplug the sensor from the device before cleaning or disinfecting it The sensor included with the Spirobank II is...

Страница 34: ...t was not performed correctly Repeat the test and follow closely the indications shown on the screen Certain spirometry and or oximetry parameters are not shown at the end of a test Personalized param...

Страница 35: ...n If the product is utilised differently from the use described in the User s Manual If any alteration adjustment modification or repair has been carried out by personnel not authorised by the manufac...

Отзывы: