Spirobank II
Rev.1.6
Page 22 of 35
User manual
DATE Format
The voice allows to choose one of the following option:
day
month year
month day
year
year
month day
Select the desired format with
or
and press
; the selection will be saved automatically and the device will return to the service
menu.
UNIT format
The voice allows to choose one of the following option:
Imperial
(in,lb)
Metric
(cm kg
)
Select the desired format with
or
and press
; the selection will be saved automatically and the device will return to the service
menu.
Info firmware
In this menu the user may view information regarding the components version
presents in the device:
•
Bluetooth version
•
Bluetooth PIN
•
Oximeter
After approximately 10 seconds the device will automatically return to the service
menu, otherwise press
.
Once all of the items in the service menu have been set it is possible to exit the
menu by pressing
.
In BASIC version the screen shows only Serial number and internal software version.
2.5.1
Reusable turbine calibration
WARNING
The turbine flow sensor does not require calibration, however regular cleaning of the turbine is necessary . The disposable
turbine is check before the application of the external packaging, for this reason it doesn’t requires a periodic calibration
. If
a calibration must be performed the following guidelines should be carefully noted.
The calibration procedure can be performed on the reusable turbine and on the disposable turbine.
Turbine calibration is performed with a calibration syringe to simulate a
FVC test for the expired parameters and a FIVC test for the inspired
parameters.
To enter the calibration function, select the “Turbine Calibration” option
from the Service Menu (as explained in paragraph 2.5). To enter the new
calibration valu
es choose the item “Modify
calibration
” in the submenu,
enter the password and insert the new calibration values. Make three
manoeuvres with a syringe as described by the screen on the device, then
Spirobank II
calculates the FVC and FIVC values.
Press
.
The screen requires to insert the volume of the syringe in use
; Spirobank
II
so calculates the correction percentage between the reference and the
calculated value. It can be possible to change the syringe volume using
and
, then press
. At this point two new correction values
are shown. Press
to apply these correction, otherwise press
to
set the factory calibration values (0%).