![background image](http://html1.mh-extra.com/html/mingtai/sun12-iii/sun12-iii_operation-manual_1801433010.webp)
8
7.2
Fill & drain water
When you install the Steam Sterilizer for the first time, after turn on the switch,
then The machine would displaying lack of the source water. Then you need to
use the measuring cup to fill distilled water from the top inlet around 4000ml for
the first time, next time do not filling the water too full.
When the machine displaying the used water full, please drain waster water out
only after sterilizing in time.
Please clean the brass filter inside the chamber, or it would effect the pressure
releasing during the sterilization. Please clean the filter in the same time with you
draining the used water.
Note: Be sure to fill distilled water .
1.
Before filling the distilled water, you must turn on the switch of the power!
2. Please fill the distilled water only after finish working!
7.3 Function description
Vacuum sterilizer divided into the following process: preheating process;the
vacuum step-up process; sterilization process; exhaust drying process; the
pressure balance process.
7.3.1 preheating process:
Select the program firstly, then close the door, press start button to enter the
preheating process,electrify heated steam generator heater reaches 100 ℃,
electrify heated the chamber wall heater to set temperature 50 ℃.