4.3.Calibration
32
Calibration
4
3. Tap [YES].
The weighing value starts to be read.
POINT
- To cancel [ZERO ADJUSTMENT], tap
[NO] then tap [BACK] or [TOP].
If [CALIBRATION 1/2] is displayed after
blinks
on the load display section, [ZERO ADJUSTMENT]
is completed.
(2) Enter the numerical value of the output voltage
Enter the output voltage value when the load is zero to register it as the zero point of the load
cell.
The accuracy of zero adjustment by entering output voltage values is about 1/1 000. When
precise accuracy more than 1/1 000 is required, perform zero adjustment by reading
weighing values (initial load values).
1) Display the screen to register [ZERO POINT mV/V]
1. Tap [ZERO POINT mV/V] on [CALIBRATION
1/2].
The input value at the time of adjustment before
shipment or previous calibration appears under
[ZERO POINT mV/V].
The screen to set output voltage values will be
displayed.
2) Set the output voltage value
1. Enter the output voltage value to register.
Unit: mV/V
Allowable range: -2.500 00 to 2.500 00
POINT
- Refer to “3.5. Input of numerical values”
for more information about how to enter
numerical values.