When the wiring, be sure to the instrument power supply is OFF.
・
Do not supply the AC power until complete the installation. This instrument does not have
power switch (ON/OFF).
・
Do not fell or make a strong impact on this instrument rear pannel terminal block because it is
made of resin.
・
Striped electrical cable tip length is 9 mm.
・
Cables which connecting this instrument isolate from noise sources, for example, power supply
line and I/O for control’s as much as possible.
・
Be sure to connect the ground wire must be D single ground. Do not common the
ground with a kind of power supply.
・
For CC
−
link cable connection, use twist pear cable wire with shield(Cable for CC
−
link) and
connect the shield in terminal block’s SLD terminal or F.G.terminal.
●
Connect the termination to the CC
−
Link connector to electrical
termination which is far from PLC as possible.
●
Use the connecting cable for CC
−
Link.