4.5 Ordering information
Model
Max. capacity
E
max
Type
PR 6211/31
30 kg
..D1
PR 6211/51
50 kg
..D1
PR 6211/12
100 kg
..D1
PR 6211/22
200 kg
..D1
PR 6211/32
300 kg
..D1
4.6 Technical data
Designation
Description
Abbr.
D1
Unit
Accuracy class
0.05
% E
max
Minimum dead load
lowest limit of speciied measuring range
E
min
0
% E
max
Maximum capacity
highest limit of speciied measuring range E
max
See Chapter
Safe load limit
maximum load without irreversible damage E
lim
500
kg
Destructive load
danger of mechanical destruction
E
d
1000
kg
minimum load cell scale interval, v
min
=
E
max
/Y
% E
max
= 50
300 kg
Y
3333
Minimum LC veriica-
tion
% E
max
= 30 kg
Y
2500
Rated output
relative output at maximum capacity
C
n
2
mV/V
Tolerance on rated
output
permissible deviation from rated output C
n
d
c
<0.25
% C
n
Zero output signal
load cell output signal under unloaded con-
dition
S
min
0
2.0
% C
n
Repeatability
max. change in load cell output for repea-
ted loading
ɛ
R
<0.01
% C
n
Creep
max. change of output signal at E
max
du-
ring 30 minutes
d
cr
<0.03
% C
n
Non-linearity
deviation from best straight line through
zero
d
Lin
<0.05
% C
n
Hysteresis
max. diference in LC output between loa-
ding and unloading
d
hy
<0.05
% C
n
Temperature efect
on Smin
max. change of S
min
per 10 K in B
T
TK
Smin
<0.042
% C
n
/10 K
Temperature efect
on C
max. change of C per 10 K in B
T
TK
C
<0.03
% C
n
/10 K
Input impedance
between supply terminals
R
LC
1300 ±12
Ω
4 Speciications
Compact Compression Load Cell PR 6211
Minebea Intec
EN-11