When a fuse has tripped:

-

Disconnect the junction box from all power sources.

-

Determine and eliminate the cause of the error.

-

Replace the fuse.
Make sure that fuses are replaced only using fuses of the same type, see Chapter

3.2

.

-

Reconnect the junction box to all power sources.

4.5 Cable gland

The cables have to be fed into the device via glands to ensure leak-tightness. Cable

diameters of 5…9.5 mm are suitable for gland M16.
The cable wires are connected to the terminals inside the device.

NOTICE

Property damage is possible.

If a cable gland is not used, it must be sealed with a supplied locking pin.

NOTICE

Property damage is possible.

Regularly check the itted cable gland for tightness and re-tighten it, if necessary.

4.6 Installation of a cable

NOTICE

Material damage is possible.

Do not guide the screen (4) into the device!

The cable shield (4) must be connected in the metal sleeve (6) of the cable gland.
Before, during and after installation, make sure that the sealing ring is seated

correctly.

4 Installation and connection information

Cable junction box for digital load cells PR 6024/68S

Minebea Intec

EN-13

Содержание PR 6024/68S

Страница 1: ... load cells PR 6024 68S Installation manual Translation of original installation manual 9499 053 46900 Edition 3 5 0 01 17 2020 Minebea Intec GmbH Meiendorfer Str 205 A 22145 Hamburg Germany Phone 49 40 67960 303 Fax 49 40 67960 383 ...

Страница 2: ...liied personnel In correspondence concerning this product the type name and release number serial number as well as all license numbers relating to the product have to be cited Note This document is partially protected by copyright It may not be changed or copied and it may not be used without purchasing or written permission from the copyright owner Minebea Intec The use of this product constitut...

Страница 3: ...peciications 7 3 1 Equipment supplied 7 3 2 Technical Data 8 3 3 Electromagnetic Compatibility EMC 9 3 4 Possible marking for the Ex area 9 3 5 Dimensions 10 4 Installation and connection information 11 4 1 General information 11 4 2 Terminals and jumpers 11 4 3 LED 12 4 4 Fuses 12 4 5 Cable gland 13 4 6 Installation of a cable 13 5 Cable connections 15 5 1 General information 15 5 2 Data and supp...

Страница 4: ...6 4 Cleaning 19 7 Disposal 20 8 Appendix 21 8 1 Certiicates safety instructions 21 Cable junction box for digital load cells PR 6024 68S Table of contents EN 2 Minebea Intec ...

Страница 5: ...sures for preventing hazards are not followed DANGER Warning of personal injury DANGER indicates death or severe irreversible personal injury which will occur if the corresponding safety measures are not observed Take the corresponding safety precautions WARNING Warning of hazardous area and or personal injury WARNING indicates that death or severe irreversible injury may occur if appropriate safe...

Страница 6: ...nment may occur if appropriate safety measures are not observed Take the corresponding safety precautions Note User tips useful information and notes 1 6 Hotline Phone 49 40 67960 444 Fax 49 40 67960 474 eMail help minebea intec com Cable junction box for digital load cells PR 6024 68S 1 Introduction EN 4 Minebea Intec ...

Страница 7: ...personnel who are aware of and able to deal with the related hazards and take suitable measures for self protection The device relects the state of the art The manufacturer does not accept any liability for damage caused by third party system components or due to incorrect use of the product The use of this product signiies recognition of the stipulations listed above The following table shows the...

Страница 8: ...signs of mechanical damage 2 5 Repairs and maintenance 2 5 1 General information Repairs are subject to inspection and must be carried out at Minebea Intec In case of defect or malfunction please contact your local Minebea Intec dealer or service center for repair When returning the device for repair please include a precise and complete description of the problem Maintenance work may only be carr...

Страница 9: ... 2 Box incl electronics 3 Ex zone locking pin 6 mm 3 4 Cable gland M16 10 5 Pressure compensation element The following items are not shown 6 Drilling template 7 Quick guide 3 Speciications Cable junction box for digital load cells PR 6024 68S Minebea Intec EN 7 ...

Страница 10: ...404 X2CrNiMo17 12 2 according to EN 10088 3 AISI 316L JIS SUS316 Surface of the junction box 2R according to DIN EN 10088 2 cold rolled bright annealed lat blank relective Net weight 2 1 kg Shipping weight 2 3 kg Service temperature area 30 C 80 C Service temperature area in explosion prone area 20 C 60 C Storage temperature range 30 C 80 C Cable screw connections Metal Ex EMC cable screw connecti...

Страница 11: ...50 µs EN 61000 4 5 1 kV Signal and control lines Conducted disturbances by high fre quency coupling 0 15 80 MHz EN 61000 4 6 10 V 3 4 Possible marking for the Ex area Zone Marking Certiicate No 2 and 22 II 3G Ex nA IIC T5 Gc II 3D Ex tc IIIC T90 C Dc SWT 12ATEX004 X NOTICE Installation in Ex Area The Ex safety instructions in the appendices must be observed when installing in the Ex area 3 Speciic...

Страница 12: ...3 5 Dimensions all dimensions in mm Cable junction box for digital load cells PR 6024 68S 3 Speciications EN 10 Minebea Intec ...

Страница 13: ...ts LC1 8 supply voltage load cells Red Terminal contact PS supply voltage supply cable PR 6124 xxP for the PR 6024 62S power supply Terminal contacts LC1 8 supply voltage load cells Blue Terminal contact PS supply voltage supply cable PR 6124 xxP for the PR 6024 62S power supply Terminal contacts LC1 8 RS 485 TxB data transmission B green Terminal contact IND RS 485 TxB data cable PR 6124 xxD for ...

Страница 14: ...voltage UDC 24 V 4 4 Fuses The power supply is protected by a time delayed 1 0 A fuse 2 in a fuse holder with protection against contact Cable junction box for digital load cells PR 6024 68S 4 Installation and connection information EN 12 Minebea Intec ...

Страница 15: ... to the terminals inside the device NOTICE Property damage is possible If a cable gland is not used it must be sealed with a supplied locking pin NOTICE Property damage is possible Regularly check the itted cable gland for tightness and re tighten it if necessary 4 6 Installation of a cable NOTICE Material damage is possible Do not guide the screen 4 into the device The cable shield 4 must be conn...

Страница 16: ...cable 2 3 Guide the cable 5 through the gland 6 4 Fold the cable shield 4 over the lower part of the terminal insert 3 approx 10 mm 5 Connect the cable conductors 6 Tighten the sleeve screw cap 1 7 Secure the gland 6 including the o ring 7 using the counter nut in the housing Cable junction box for digital load cells PR 6024 68S 4 Installation and connection information EN 14 Minebea Intec ...

Страница 17: ...lls via the PR 5510 04 interface and the PR 6024 68S junction box 5 2 Data and supply cables The PR 6124 xxD data cable must be used to connect the PR 6024 68S cable junction box to the device max length 300 m The PR 6124 xxP power cable must be used to connect the PR 6024 62S power supply to the PR 6024 68S cable junction box max length 200 m 5 3 Cable connections Note All components are only sho...

Страница 18: ...Connecting 8 load cells to the PR 5510 04 interface card Cable junction box for digital load cells PR 6024 68S 5 Cable connections EN 16 Minebea Intec ...

Страница 19: ...Connecting 12 load cells to the PR 5510 04 interface card 5 Cable connections Cable junction box for digital load cells PR 6024 68S Minebea Intec EN 17 ...

Страница 20: ...ting the protective grounding conductor is not enough If the installation is not carried out according to our instructions this voids the warranty In particular the entire installation including the power supply must be suiciently protected against lightning In Ex areas equipotential bonding must be established to avoid any compensatory currents between the individual conducting system components ...

Страница 21: ...ldering work on the device is neither required nor permitted 6 4 Cleaning NOTICE Property damage caused by unsuitable cleaning utensils agents Damage to the device Prevent moisture from penetrating the interior Do not use aggressive cleaning agents solvents or similar agents For use in the food industry use a cleaning agent suitable for that particular working environment Use soft sponges brushes ...

Страница 22: ...nicipal waste so that it can then be recycled Before disposing of or scrapping the product any batteries should be removed and taken to a suitable collection point Please see our T Cs for further information Service addresses for repairs are listed in the product information supplied with the product and on our website www minebea intec com We reserve the right not to accept products that are cont...

Страница 23: ...iption Document no 1 EU Declaration of Conformity MEU17054 2 Declaration of Conformity MDC17005 3 Manufacturer s Certiicate SWT 12 ATEX004 X 4 Safety instructions 65835 790 16 8 Appendix Cable junction box for digital load cells PR 6024 68S Minebea Intec EN 21 ...

Страница 24: ...1 6 ...

Страница 25: ...2 6 ...

Страница 26: ... Herstellers 2 1 und seines Bevollmächtigten 2 2 3 Die alleinige Verantwortung für die Ausstellung dieser Konformitätserklärung trägt der Hersteller 4 Gegenstände der Erklärung 5 Die oben beschriebenen Gegenstände der Erklärung erfüllen die einschlägigen Harmonisierungsrechtsvorschriften der Union 6 Angabe der einschlägigen harmonisierten Normen oder der anderen technischen Spezifikationen die der...

Страница 27: ... proizvodu Aέ1 τгnačavanje magyar hu italiano it δatvių kalba lt εegfelelősцgi nвilatkoгat Dichiarazione di conformità Atitikties deklaracija 1 Termékmodell termékszám kizárólag az alábbi projektszámhoz érvényes 2 A gyártó 2 1 vagy adott esetben meghatalmaгott kцpviselőjцnek 2έ2 neve цs címe 3έ Eгt a megfelelősцgi nвilatkoгatot a gвпrtя kiгпrяlagos felelőssцge mellett aНjпk kiέ 4 A nyilatkozat tár...

Страница 28: ...tgevoerd en het certificaat y verstrekt dat relevant is voor z A Aanvullende informatie over A 1 Markering polski pl português pt romсnă ro ϊeklaraМja гgoНnośМi Declaração de conformidade ϊeМlarație Нe Мonformitate 1έ εoНel proНuktu ή numer proНuktu ή аażnв авłąМгnie Нla projektu o numerгeμ 2 Nazwa i adres producenta 2 1 oraz jego upoаażnionego przedstawiciela 2 2 3έ σiniejsгa НeklaraМja гgoНnośМi...

Страница 29: ...organ а je iгveНel б in izdal certifikat y pomemben za z A Dodatne informacije o A 1 Oznaka 1 Tuotemalli tuotenumero koskee vain projektinumeroa 2 Valmistajan 2 1 ja valtuutetun edustajan 2 2 nimi ja osoite 3 Tämä vaatimustenmukaisuusvakuutus on annettu valmistajan yksinomaisella vastuulla 4 Vakuutuksen kohde kohteet 5 Edellä kuvattu kuvatut vakuutuksen kohde kohteet on ovat asiaa koskevan unionin...

Страница 30: ......

Страница 31: ......

Страница 32: ......

Страница 33: ...den Fall der Patent Gebrauchsmuster oder Geschmacksmustereintragung vorbehalten Teil Dok Nr Part Doc No 1 Main connection Explosion risk area Zone 2 or Zone 22 Explosion safe area Power supply PR6024 62S Ta 20 C 50 C 2 Data transfer Indicator only for data transfer there is no supply voltage going from the connected device to the cable junction box This device is also suitable for installation in ...

Страница 34: ...re used in the manner in which they are laid and their relevant mechanical loads Cables sourced from other manufacturers are the operator s responsibility 7 The ATEX certified cable glands 7 and 8 are only suitable for fixed installation when used in potentially explosive atmospheres meaning that the equipment installer has to fit the cable behind the cable gland e g using a cable clip These cable...

Страница 35: ......

Страница 36: ...Published by Minebea Intec GmbH Meiendorfer Strasse 205 A 22145 Hamburg Germany Phone 49 40 67960 303 Email info minebea intec com www minebea intec com ...

Отзывы: