background image

3.3.2.2.1 Support bracket

The remote display with table-top housing has a support bracket with drilled holes at

various intervals.
The table-top housing is secured in the desired position using handwheel bolts (not

shown).

Front view

Side view

All dimensions in mm

All dimensions in mm

Bottom side

All dimensions in mm

3 Remote display

Remote display PR 5110/60, ../70

EN-11

Minebea Intec

Содержание PR 5110

Страница 1: ...70 Instrument manual Translation of the original instrument manual 9499 050 51600 Edition 3 1 0 02 07 2019 Release 1 00 Minebea Intec GmbH Meiendorfer Str 205 A 22145 Hamburg Germany Phone 49 40 6796...

Страница 2: ...ied personnel In correspondence concerning this product the type name and release number serial number as well as all license numbers relating to the product have to be cited Note This document is par...

Страница 3: ...7 2 4 5 Failure and excessive stress 7 2 4 6 Important note 7 2 4 7 Repairs and maintenance 7 3 Remote display 8 3 1 General notes 8 3 2 Overview of the instrument 8 3 2 1 Communication 8 3 3 Housing...

Страница 4: ...ance 27 6 2 Repairs 27 6 3 Soldering work 27 6 4 Cleaning 27 7 Disposal 28 8 Technical data 29 8 1 Equipment supplied 29 8 2 Decoding the serial number 29 8 3 Possible marking for the Ex area 29 8 4 G...

Страница 5: ...ed when measures for preventing hazards are not followed DANGER Warning of personal injury DANGER indicates death or severe irreversible personal injury which will occur if the corresponding safety me...

Страница 6: ...the environment may occur if appropriate safety measures are not observed Take the corresponding safety precautions Note User tips useful information and notes 1 6 Hotline Phone 49 40 67960 444 Fax 4...

Страница 7: ...relects the state of the art No warranty is given that the product is free of faults especially not in conjunction with third party software and hardware components required for operation The manufac...

Страница 8: ...may be exposed Please note that capacitors in the device may still be charged even after disconnecting the device from all voltage sources Disconnecting the device from the power supply This device c...

Страница 9: ...ransport 2 4 6 Important note Make sure that the construction of the device is not altered to the detriment of safety In particular leakage paths air gaps of live parts and insulating layers must not...

Страница 10: ...parity bit straight parity 1 start bit and 1 stop bit If this data format including correct parity is not maintained the complete data telegram is considered not received and shown on the remote disp...

Страница 11: ...3 2 Dimensions 3 3 2 1 Control cabinet housing The keypad and the display form one unit with the front A rectangular cut out is required for the installation Cable connections are made at the back of...

Страница 12: ...nd the display form one unit with the front The cables are fed through entry glands on the bottom of the housing and connected to the electronics Front view Side view 238 170 All dimensions in mm 85 3...

Страница 13: ...with drilled holes at various intervals The table top housing is secured in the desired position using handwheel bolts not shown Front view Side view All dimensions in mm All dimensions in mm Bottom s...

Страница 14: ...escription 1 Weighing range 2 Weighing point 3 Weight type plus or minus sign standstill 4 Weight value 5 Optional framing 6 Symbols mass unit Weight type plus or minus sign standstill Description Gro...

Страница 15: ...displayed weighing point in the indicator provided that the weight value is stable the device is not in error status The key can be deactivated in parameter OP 016 The currently displayed weighing po...

Страница 16: ...set up Previous digit value selection Select weighing point Cursor to the left Exit Exit from current menu continue operation at next higher level Cursor to the right OK Parameter selection conirmatio...

Страница 17: ...mplete including cables e g power cable fuse protection load cells junction box data cables console cabinet etc Follow all device installation instructions related to application safety ventilation se...

Страница 18: ...he wires must be kept as short as possible leading to the terminals The plug housings are conductive metallic and thus also part of the screen They must be screwed correctly to the back wall NOTICE Ma...

Страница 19: ...uring and after installation make sure that the sealing ring is seated correctly 4 Cut back the cable screen 3 up to 5 mm and bend back over the cable mantel 1 5 Remove the insulation of the cable str...

Страница 20: ...ousing 9 pin D Sub male connector Table top housing 5 pin plug in screw terminal Technical data Description Data Number of channels 1 Type TTY Transmission rate bit s 2 400 9 600 Number of databits 7...

Страница 21: ...reened twisted pair suchas LifYCY 3 2 0 20 Cable gauge 0 25 0 5 mm2 Cable length Max 1500 m TxE RxD TxD 2 TxE RxD TxD TxE 8 RxD 9 4 TxD 5 Com 1 1 5 4 3 2 1 6 1 2 9 pin D Sub connector male switchbox h...

Страница 22: ...al equalization line Coniguration PR 5110 for PR 5220 PR 5230 PR 5410 PR 5220 PR 5230 PR 5410 oP 10 LInE curloop oP 12 tokEn oFF oP 13 SEndModE SEnd oP 15 WPkEY SELECt Serial ports parameter Remote di...

Страница 23: ...general we recommend that after coniguration you set the switch to the closed position to prevent accidental overwriting of the parameters For legal for trade applications the LOCK switch must be seal...

Страница 24: ...V intrinsically safely fed is correct and whether all connections are made correctly 5 2 1 Automatic tests After connecting the power supply there is a display test All elements are displayed for 2 se...

Страница 25: ...2 Press key to display the parameter name If key is pressed again the irst selectable option of this parameter is displayed Press keys or to select the parameter Press key to go to the next parameter...

Страница 26: ...dental alteration and power failure 5 3 1 1 Selecting a digit position To go to the next digit press the key To go to the previous digit press the key 5 3 1 2 Changing a digit value To increase the va...

Страница 27: ...5 3 2 Procedure Factory Settings default An invalid weight is displayed without a mass unit 5 Getting started Remote display PR 5110 60 70 EN 25 Minebea Intec...

Страница 28: ...o Net A when available Weight tare set to tare A when available oP 014 Weight follow Display follows the indicator weight mode Wpkey select Weighing point selected from one or multiple indicators Pres...

Страница 29: ...x devices 6 3 Soldering work Soldering work on the device is neither required nor permitted 6 4 Cleaning NOTICE Property damage caused by unsuitable cleaning utensils agents Damage to the device Preve...

Страница 30: ...unsorted municipal waste so that it can then be recycled Before disposing of or scrapping the product any batteries should be removed and taken to a suitable collection point Please see our T Cs for...

Страница 31: ...00015 Location no 30 Hamburg Code for the year month 252 April 2010 Current number Is increment according to the year group table of Minebea Intec 8 3 Possible marking for the Ex area Zone Marking Cer...

Страница 32: ...operation Dust 0 50 C Limits for storage transport 20 70 C Moisture Limits for operation 10 90 non condensing pursuant to IEC 60068 2 Limits for storage transport 5 95 non condensing pursuant to IEC...

Страница 33: ...0 40 70 bis 80 ms Voltage luctuations EN 61000 4 11 20 Mains inputs Short term disturbances EN 61000 4 11 0 5 s Disruptions via intrinsic feed 8 5 3 Interference suppression Electromagnetic emissi on...

Страница 34: ...no Description Document no 1 EC Declaration of Conformity MEU17055 2 EC Type Examination Certiicate KEMA 10 ATEX 0090X The documents listed in the table can be found on the PR 5110 60 70 CD Remote di...

Страница 35: ......

Страница 36: ...Published by Minebea Intec GmbH Meiendorfer Strasse 205 A 22145 Hamburg Germany Phone 49 40 67960 303 Email info minebea intec com www minebea intec com...

Отзывы: