Note:
For further information, see the platform scale operating instructions.
4.5.3
PR 5500/07 analog input and output
The plug-in card for the available analog channels has 1 analog output (active) and 1
analog input.
The card is inserted in the option 1/FB and/or option 2 slot.
A maximum of 2 PR 5500/07 cards can be installed.
Technical data
Description
Data
Output
Output
1 active current output: 0/4 to 20 mA (max. 24 mA),
10 V output voltage via an external 500
Ω
resistor
Output function
Gross/Net/Display follows or application-dependent
Output range
0/4 to 20 mA, conigurable
Output resolution
Internal 16 bits = 65,536 counts,
resolution of 20,000 @ 20 mA
Output linearity error
@ 0 to 20 mA: 0.04%
@ 4 to 20 mA: 0.02%
Process Controller Maxxis 4 PR 5500