9.7.2
Stability detection time (STAB. DETECTION TIME)
The time width of stability detection is set in order to set the operative condition of
Stability detection.
A time is measured just after the change width of the load display becomes within the
efective range, and this funciton judges whether the time width is within the range or not
set before.
The set value is input with ten keys in the range from 0.0 to 9.9 in every 0.1 seconds.
The stability detection turns of when the setting is [0.0].
[1.0] is set as default.
9.7.3
Operational condition of detecting stability (OPERATION OF STABILITY)
This unit is always detecting the stability in each setting screen.
The operational condition of stability detection is selectable from [ANYTIME] or
[WEIGHING SCREEN ONLY]. [ANYTIME] is selected as default.
9.8 Zero set
The zero set function memorize the present weighing value as zero point and make the
display to zero by tapping the [ZERO] when the display value at weight display part is the
gross value, and at the same time the value is within the efective range of zero set.
9.8.1
Operative condition of zero set (OPERATION OF ZERO)
The operative condition of zero set is selectable from [IN STABLE MODE] or [ANYTIME].
The efective range is common for the zero set and the zero tracking.
[ANYTIME] is set as default.
9.8.2
Efective range of zero set (RANGE OF ZERO)
The efective range of zero set is set.
The setting is input with ten keys in the range from 0 to 30 in every 1 % with the
persentage of max. load (Max).
The zero set turns of when the setting is [0].
[5] is set as default.
Note:
-
The error of [Z. ERR ZERO] displays when the zero set function is set out of the
efective range, and zero set is not accepted.
Tap [ERR. RESET] to cancel the error on top left comer of the screen.
-
When the tare weight cancellation or the preset tare weight cancellation is set,
please forward the zero set after the clear of them.
9.9 Zero tracking
The zero tracking function corrects the zero drift, and stabilizes the zero point when the
change width of the load display is within the data width for zero tracking, and at the
same time the condition continues longer than the time ofset by zero tracking time.
Weight Controller CSD-918
9 Basic function setting display (BASIC FUNCTION)
EN-78
Minebea Intec