[Z MV]
[
✶
.
✶✶✶✶✶
] is the saved load cell output voltage.
mV/V
Saved load cell output voltage.
Set the load cell output voltage in mV/V corresponding to the measurement zero
point.
[TARE WEIGHT] [NET-GROSS WEIGHT]: The changed digit value adjusts.
[ACCUM] [PRESET TARE WEIGHT]: The changed digit is selected.
[ZERO]: The value changes to [0].
[SET]: Interrupts the setting and returns to the CCAL display
[F]: The displayed value is saved. Proceed to the next step.
mV/V
Set the load cell output voltage:
[SPAN]
▼ SPAN point calibration
Error display in zero point calibration:
[TE-L]: The display blinks for about two seconds if load cell output is –2.5
mV/V or less (refer to Chapter
[TE-H]: The display blinks for about two seconds if load cell output is 2.5
mV/V or more (refer to Chapter
).
6.3.7
SPAN calibration
[SPAN] is displayed when SPAN calibration starts.
Select the SPAN calibration method:
-
Method by weight
⇨
Press the [F] key and execute "Method by weight".
Execute SPAN calibration with a weight on the load cell.
-
Method by numeric output of load cell output
⇨
Press the [PRESET TARE WEIGHT]
key and execute the "Method by numeric input".
Execute SPAN calibration by inputting the diference between the output voltage of
the load cell and measurement zero point (Setting described in Chapter
output voltage corresponding to the mass of the weight (Setting described in
Chapter
1. When the[PRESET TARE WEIGHT] key is pressed, [Z MV] is displayed.
2. When the[F] key is pressed, [
✶
.
✶✶✶✶✶
] is displayed.
3. Press the [F] key after setting
6 Calibration
Weight indicator CSD-903
Minebea Intec
EN-41