Perform the following cleaning and disinfection for each patient: Wipe ventilator external

surface (including housing, power cord, supply gas hose), trolley, support arm, and touch 

screen for cleaning. Wipe or place them under ultraviolet radiation for disinfection. 

 

y

Perform the following cleaning and disinfection for each patient or every week:

Immerse expiration valve assembly (except membrane), expiration valve membrane 

and patient tube (including water trap, Y piece, adapter) in detergent for cleaning. 

Immerse or autoclave them for disinfection.

 

For cleaning and disinfection of 

mainstream CO

2

 sensor, SpO

2

 sensor, SpO

2

 sensor cable, nebulizer and humidifier, 

refer to the instructions provided by the vendors.

 

y

Perform the following cleaning and disinfection as necessary: Immerse inspiration

safety valve assembly in detergent for cleaning. Immerse or autoclave it for 

disinfection.

 

y

Perform the following cleaning and disinfection every 4 weeks or as necessary: Immerse

the fan dust filter, main unit air outlet dust filter, and air intake dust filter in detergent 

for cleaning and disinfection.

Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Floor 1-4, Mindray Building, Keji 12th Road South, High-tech Industrial Park, Nanshan, Shenzhen, P. R. China
Tel:0755 81888998    

Fax:0755 26582680

www.mindray.com

Cleaning and Disinfection:

 

y

Annually or as necessary, check the inspiration safety valve membrane. Contact 

your 

service personnel 

for replacement if necessary.

 

y

Each year, per 5000 hours or as necessary, replace air intake HEPA filter. Check the

expiration valve membrane and contact us for replacement if necessary. If the O

2

 Cell 

damages, please replace one. Please contact us for preventive maintenance. 

 

y

Replace clock module battery every six years or as necessary. Contact us for

replacement if necessary.

y

 

Please contact us if the preventive and maintenance buttons 

 

are shown on the 

system check interface or circuit test interface.

The table below lists the cleaning and disinfecting agents and autoclaving process that may 
be used on the ventilator.

Name

Type

Ethanol (75%)

Moderately efficient disinfectant

Islpropanol (70%)

Moderately efficient disinfectant

Glutaraldehyde (2%)

High-efficient disinfectant

Ortho-phthalaldehyde (e.g. Cidex®OPA

High-efficient disinfectant 

Soap water (pH value of 7.0~10.5) 

Rinsing agent

Clean water 

Rinsing agent

Steam autoclave (134

 recommended)

Highly efficient disinfection

Basic Operation, Daily Maintenance and Disinfection Guidelines

-Page 4 of 4-

0 4 6 - 0 0 9 9 8 3 - 0 0 ( 4 . 0 )

Содержание SV600

Страница 1: ... are no absolute contraindications for this product For some special diseases however some necessary treatments shall be taken for ventilator mechanical ventilation or special ventilation modes shall be adopted to prevent possible patient injury This guideline briefs the ventilator its basic operation daily maintenance cleaning and disinfection For more information please refer to the Operator s M...

Страница 2: ...Neonatal flow sensor connector N Module slot a b c d f e a Mode and Patient size area b Information display area c Icon area d Functional key area e Vent mode and Parameter Setting area f Waveforms Parameter Spirometry Dynamic Lung Trend Display Area Right View Left View Main Screen Basic Operation Daily Maintenance and Disinfection Guidelines Page 2 of 4 ...

Страница 3: ...e y y Perform the following maintenance several times a day or as necessary check the breathing tubes and water traps for water build up Empty water build up if there is Inspect the parts for damage Replace as necessary y y Perform the following maintenance each patient or as necessary Perform pressure and flow zeroing system check and flow sensor calibration Replace with disinfected parts or new ...

Страница 4: ...oor 1 4 Mindray Building Keji 12th Road South High tech Industrial Park Nanshan Shenzhen P R China Tel 0755 81888998 Fax 0755 26582680 www mindray com Cleaning and Disinfection y y Annually or as necessary check the inspiration safety valve membrane Contact your service personnel for replacement if necessary y y Each year per 5000 hours or as necessary replace air intake HEPA filter Check the expi...

Отзывы: