background image

1.

Connect the ventilator and humidifier to the AC power supply. Connect the gas supply.

2.

Start the humidifier and set the relevant parameters by following the humidifier 

instructions for use (check if the water level of the humidifier complies with the 

requirement before starting the humidifier).

3.

When the ventilator is turned on, it runs self-test automatically.

4.

Connect the breathing tubes, and select to run (recommended) or skipthe system check.

5.

Once system check is finished, the ventilator starts to work in the set mode. For new patients, 

you need to set the IBW, Patient Type (adult, pediatric or neonate), Gender, and Vent Type 

(invasive or non-invasive).

6.

Connect the test lung, and check the status of the test lung. If the test lung is in good 

condition, you can disconnect the test lung and connect the ventilator to the patient 

(recommended).

7.

Connect the patient and breathing tubes, and start ventilation. Keep the patient under 

close monitoring for a period and pay attention to the vital signs of the patient and 

ventilation status of the ventilator.

8.

Keep observing the parameters and alarms during ventilation. Reconfigure the ventilation 

mode or parameters if necessary.

9.

After use, turn off the ventilator and humidifier in sequence.

Basic Operation:

Routine Maintenance:

 

y

Before each use or after continuous use of two weeks, perform system check. Check

the breathing system resistance and leakage.

 

y

Perform the following maintenance several times a day or as necessary: check the

breathing tubes and water traps for water build-up. Empty water build-up if there is. 

Inspect the parts for damage. Replace as necessary.

 

y

Perform the following maintenance each patient or as necessary: Perform pressure and 

flow zeroing, system check and flow sensor calibration. Replace with disinfected parts or 

new disposable parts.

 

y

Perform the following maintenance daily or as necessary (recommended):Clean the

external surfaces and calibrate the O

2

 sensor. 

Replace the O

2

 sensor if it is damaged. 

When the patient’s exhaled gas may contaminate the inspiratory safety valve 

assembly, it is necessary to replace with disinfected inspiratory safety valve and 

membrane.

Replace the expiration valve if it is damaged.Calibrate the CO

2

 module 

when the CO

2

 measured value has a great deviation. Calibrate the touch screen if its 

function is degraded. Check water traps at air supply inlet. Empty water build-up if 

there is. Please contact your service personnel if any cracks and leakage are found on 

the water traps.

 

y

Perform the following maintenance monthly or as necessary: Check the air intake

dust filter, fan dust filter  and main unit air outlet dust filter for dust build-up. Clean or 

replace as necessary.

 

y

When cleaning or installation, check the parts for damage and replace as necessary.

 

y

Check lithium battery charging and discharging every 6 months and replace every

three years. Please contact 

your service personnel

 for replacement.

Basic Operation, Daily Maintenance and Disinfection Guidelines

-Page 3 of 4-

Содержание SV600

Страница 1: ... are no absolute contraindications for this product For some special diseases however some necessary treatments shall be taken for ventilator mechanical ventilation or special ventilation modes shall be adopted to prevent possible patient injury This guideline briefs the ventilator its basic operation daily maintenance cleaning and disinfection For more information please refer to the Operator s M...

Страница 2: ...Neonatal flow sensor connector N Module slot a b c d f e a Mode and Patient size area b Information display area c Icon area d Functional key area e Vent mode and Parameter Setting area f Waveforms Parameter Spirometry Dynamic Lung Trend Display Area Right View Left View Main Screen Basic Operation Daily Maintenance and Disinfection Guidelines Page 2 of 4 ...

Страница 3: ...e y y Perform the following maintenance several times a day or as necessary check the breathing tubes and water traps for water build up Empty water build up if there is Inspect the parts for damage Replace as necessary y y Perform the following maintenance each patient or as necessary Perform pressure and flow zeroing system check and flow sensor calibration Replace with disinfected parts or new ...

Страница 4: ...oor 1 4 Mindray Building Keji 12th Road South High tech Industrial Park Nanshan Shenzhen P R China Tel 0755 81888998 Fax 0755 26582680 www mindray com Cleaning and Disinfection y y Annually or as necessary check the inspiration safety valve membrane Contact your service personnel for replacement if necessary y y Each year per 5000 hours or as necessary replace air intake HEPA filter Check the expi...

Отзывы: