P/N: 046-003757-00 Rev A
Quick Reference Guide
11
2.
Continue MPM
to continue with the patient data and settings in the
original MPM and clears all patient data in the new monitor.
All patient
data in the original MPM will copy to the new monitor.
3.
New Patient
to delete all patient data in the original MPM and all
patient data in the new monitor. Default settings for the patient
category are applied to the new patient.
All patient data in the
original MPM and in the new monitor will be deleted.
4.
Same Patient
to merge the patient data from the original MPM with
the patient data in the new monitor.
Panorama® Central Station Interface:
Transferring patient data from one
DPM 6 or DPM 7 bedside monitor to another DPM 6 or DPM 7 bedside monitor
will not transfer data within the Panorama® Central Station. Transferring data
within a Panorama® Central Station or to a different Panorama® Central Station
may affect the DPM 6 or DPM 7 bedside monitor.
HEMODYNAMIC MONITORING
Initiate A Cardiac Output Measurement
Cardiac Output (CO) is the amount of blood ejected from the left ventricle
each minute, expressed in liters per minute (l/min). Cardiac Index (CI) is the
Cardiac Output divided by the patient's body surface area.
To initiate a Cardiac Output measurement on the DPM 6 or DPM 7 bedside
monitor:
a. Connect the C.O. cable to the C.O. connector on the monitor
interconnecting the C.O. Module, PA Catheter, and injectate.
b. Select
C.O. Setup
menu by selecting C.O. tile,
(quick key), or from
Parameters
menu (
or
Main Menu
quick keys)
c. Verify the computation constant or adjust the computation constant in
the
Setup
menu
d. Inject solution when
“Inject Now”
message is displayed (Auto measure
mode is enabled in
Setup
menu)
Or
d. Select
Start
and inject solution within 4 seconds to perform a cardiac
output measurement (Manual measure mode is enabled in
Setup
menu)