3-12
BeneVision N22/BeneVision N19 Patient Monitor Service Manual
3.4.2
Resp Performance Test
Tools required:
Medsim300B patient simulator or other equivalent simulator
1.
Connect the patient simulator to the module using a non ESU-proof cable and set lead II as the respiration lead.
2.
Configure the simulator as follows: lead II as the respiration lead, base impedance line as 500 Ω; delta impedance
as 1 Ω, respiration rate as 20 rpm.
3.
Verify that the Resp wave is displayed without any distortion and the displayed Resp value is within 20±1 rpm.
3.4.3
SpO
2
Test
Tools required:
None.
1.
Connect SpO
2
sensor to the SpO
2
connector of the monitor. Set Patient Category
to Adult and PR Source to SpO2
on the monitor.
2.
Apply the SpO
2
sensor to the ring finger of a healthy person.
3.
Check the Pleth wave and PR reading on the screen and make sure that the displayed SpO
2
is within 95%
and100%.
4.
Remove the SpO
2
sensor from your finger and make sure that an alarm of SpO
2
Sensor Off is triggered.
Measurement accuracy verification:
The SpO
2
accuracy of the MPM module has been verified in human experiments by comparing with arterial blood
sample reference measured with a CO-oximeter. Pulse oximeter measurements are statistically distributed and about
two-thirds of the measurements are expected to come within the specified accuracy range compared to CO-oximeter
measurements.
NOTE
A simulator cannot be used to assess the accuracy of a pulse oximeter monitor or a SpO
2
sensor. Instead, it can only verify that whether the monitor is functional. The accuracy of a
pulse oximeter monitor or a SpO
2
sensor needs to be verified by clinical data.
3.4.4
NIBP Tests
See section
3.2.23.2.2NIBP Tests
.
Содержание BeneVision N19
Страница 1: ...BeneVision N22 BeneVision N19 Patient Monitor Service Manual...
Страница 2: ......
Страница 50: ...3 24 BeneVision N22 BeneVision N19 Patient Monitor Service Manual FOR YOUR NOTES...
Страница 60: ...4 10 BeneVision N22 BeneVision N19 Patient Monitor Service Manual FOR YOUR NOTES...
Страница 84: ...6 20 BeneVision N22 BeneVision N19 Patient Monitor Service Manual FOR YOUR NOTES...
Страница 100: ...7 16 BeneVision N22 BeneVision N19 Patient Monitor Service Manual FOR YOUR NOTES...
Страница 104: ...A 4 BeneVision N22 BeneVision N19 Patient Monitor Service Manual FOR YOUR NOTES...
Страница 105: ...046 011262 00 3 0...