18
BeneVision N22/N19/N17/N15/N12
Quick Reference Guide
To adjust the alarm properties of parameters on the N Series bedside monitor:
a. Select the
Alarm Setup
quick key.
Or
a. Select the
Main Menu
quick key, and
select
Limits
from the
Alarm
column.
b. Select parameter and select
•
On/Off
– to enable or disable parameter's alarm responses.
•
Select
High
or
Low
limit and adjust value using the drop down value list
•
Priority
– to adjust the audio and visual alarm indicators. Level options
include High (Red), Medium (Yellow), Low (Yellow) and Prompt (audio
absent).
•
Alarm Output
– to activate an automatic recording/printing when the
alarm
occurs.
c. Select
the
Main Menu
quick key or
to return to the main display.
Changing Arrhythmia Alarm Settings
To adjust a patient’s arrhythmia alarm on the N Series bedside monitor:
a. Select
Alarm Setup
quick key,
ECG
waveform/numeric tile or
Main
Menu
quick key.
b. Select
Arrhythmia
tab.
c. Scroll and select individual arrhythmia.
d.
Select
•
On/Off
– to enable or disable arrhythmia's alarm responses.
•
Priority
– to adjust the audio and visual alarm indicators. Level options
include High (Red), Medium (Yellow), Low (Yellow) and Prompt (audio
absent).
Note:
If alarm is locked, alarm cannot be turned off or priority altered.
•
Alarm Output
– to activate an automatic recording when the alarm
occurs.
e. Select the
Main Dialog
quick key or
to return to the main display.
To adjust a patient arrhythmia threshold setting on the N Series bedside monitor: