BeneLink Module Operator’s Manual
9-9
9.3.5
Setting the Serial Ports of Other External Devices
Brand
Model
Protocol
Baud
Rate
Data
Bits
Stop
Bits
Parity
Remark
Masimo/ pulse
CO-oximeter
Radical 7
IAP
9600
8
1
/
Method 1: Connecting Radical 7 via Docking
Station: Main Menu
→
Device Settings
→
Device output
→
Serial : IAP
Connecting Radical 7 via Root:
Step1: Main Menu
→
Device Settings
→
Device Output
→
IAP
Step2: Main Menu
→
Device Settings
→
Access Control(PW:6274) Output
→
set USB
Port to 1 or 2,, baudrate to 9600
Organon/ NMT
monitoring
device
TOF-Watch
SX
/
19200
8
1
None
No Need to set.
The serial interface adapter with cable from
Organon is required.