BeneLink Module Operator’s Manual
2-7
2.5
Configuring the ID Adapter
To configure the ID adapter to work with your external device, follow this procedure:
1.
Select [
Update ID module
]:
For BeneVision N Series monitor, select the [
Main Menu
] quick key
→
from the [
System
] column select [
Maintenance
]
→
input the factory
maintenance password
→
select [
Ok
]
→
select the [
Factory
Maintenance
] tab
→
select the [
Setup
] tab
→
select [
Update ID
module
].
For Passport 12m/17m, select [
Main Menu
]
→
[
Factory
Maintenance>>
]
→
input the factory maintenance password
→
[
Update ID module
].
2.
Set [
Benelink Module Port
] and [
ID
].
The setting of [
Benelink Module Port
] shall be consistent with the
port the RJ45 connecting cable is connected to.
You must connect the RJ45 connecting cable to the selected port when
configuring the ID adapter. Otherwise, ID adapter configuration will
fail.
Set [
ID
] to configure a new ID to the ID adapter. Refer to Column “ID for
ID Adapter” in the tables above to obtain the ID information for each
external device.
CAUTION
ID adapter setup should only be conducted by Mindray authorized
personnel.