BeneLink Module Operator’s Manual
4-65
Monitor output: alarms from Hamilton Galileo (Polling Protocol) Ventilator
Priority
Label
Medium
Medium
Technical
alarms
If the monitor shows the alarm [
Medium Technical
alarms
], the external device may have the alarm:
Gas Supply
Low
Low
Technical
alarms
If the monitor shows the alarm [
Low Technical
alarms
], the external device may have the following
alarms:
Operator
General Alarm
4.15
MAQUET SERVO-i/SERVO-s
4.15.1
Output Signals – Parameters
Monitor output: parameters from MAQUET SERVO-i/SERVO-s Ventilator
Label
Description
Trend, record, print
Monitoring Parameters
O2%
Oxygen concentration
Yes
PEEP
Positive end-expiratory pressure
Yes
Ppeak
Peak pressure
Yes
Pplat
Plateau pressure
Yes
Pmean
Mean pressure
Yes
VTe
Expiratory tidal volume
Yes
VTi
Inspired tidal volume
Yes
MVspn
Spontaneous breathed minute volume
Yes
MVe
Expiratory minute volume
Yes
MVi
Inspiratory mimute volume
Yes
ftot
Total respiratory rate
Yes
fspn
Spontaneous respiratory rate
Yes
I:E
Inspiratory time:Expiratory time ratio
No
Содержание BeneLink
Страница 1: ...BeneLink Module Operator s Manual...
Страница 2: ......
Страница 16: ...FOR YOUR NOTES...
Страница 20: ...1 4 BeneLink Module Operator s Manual FOR YOUR NOTES...
Страница 168: ...FOR YOUR NOTES...
Страница 176: ...6 4 BeneLink Module Operator s Manual FOR YOUR NOTES...
Страница 182: ...8 2 BeneLink Module Operator s Manual FOR YOUR NOTES...
Страница 227: ...P N 046 011948 00 3 0...