Accutorr Plus™ Operating Instructions
4 - 1
4.0
User Maintenance
4.1
Introduction
This section of the manual outlines routine maintenance that should be performed by the user.
The Accutorr Plus is designed for stable operation over long periods of time and under
normal circumstances should not require technical maintenance beyond that described in this
section. However, it is recommended that routine maintenance calibration and safety checks
be performed at least once a year, or more often as required by local statutory or hospital
administration practice.
4.2
Care and Cleaning Of Monitor
The monitor enclosure may be cleaned with a mild soap and water solution or ammoniated
window cleaner. Apply cleaning solution to the cloth, not directly onto the monitor. DO NOT
apply large amounts of liquid. DO NOT use abrasive cleaning agents or organic solvents.
Remove dust and dirt particles with a soft sponge moistened with cleaner solution; or a fine,
soft-hair brush. To prevent scratches DO NOT use abrasive cleaning materials. Fingerprints
and stains may be removed by using a liquid lens cleaner and a soft cloth. DO NOT wipe a
dry screen or use alcohol or chlorinated hydrocarbon solvents.
4.2.1
Decontamination of the Accutorr Plus
WARNING: Perform this process with the unit powered down and
Decontamination of a unit that has come in contact with a biological material can be
performed using LpH SE Germicidal detergent. Apply a small amount of detergent to a
disposable wipe (paper based) and wipe down the outside of the unit. Discard the wipe
appropriately. After waiting 10 minutes, use a clean dry wipe to dry the unit.