INSTRUCTIONS FOR USE SUPPORTS AND STANDS
2.5 OPERATION AND STORAGE CONDITIONS
Different environmental conditions apply for the operation of the device and its intermediate storage.
2.6 CLEANING AND DISINFECTION
3
USE OF THE LIGHTWEIGHT SUPPORT
USE OF THE LIGHTWEIGHT SUPPORT
CLEANING
WARNING – RISK OF INFECTION AND CONTAMINATION FOR PATIENTS
Parts of the support arm system and the adaptations are made of plastic. Solvents can etch plastics. Strong
acids, alkalis and media with more than 60% alcohol can cause embrittlement of the plastics.
Damaged parts can fall into open wounds.
If cleaning fl uid penetrates into the support system and the adaptations, excess solution could get into open
wounds.
DISINFECTION
WARNING – HEALTH HAZARD
Disinfectants can contain harmful substances which, in contact with skin and eyes, may cause injury or
damage to the respiratory system through inhalation. Adhere to protective measures:
•
Observe hygiene guidelines.
•
Observe the disinfectant manufacturer‘s instructions.
•
Perform surface disinfection each working day and in case of contamination (see also section “Cleaning
and disinfection”).
NOTE – QUALIFICATION OF PERSONNEL
The device shall be operated only by medical professionals in the hospital and in the doctor‘s offi ce.
ENVIRONMENTAL CONDITIONS FOR THE STORAGE AND TRANSPORT
The following storage conditions apply for up to 15 weeks after the date of delivery:
Ambient temperature:
-25 °C to 70 °C
Relative humidity (non-condensing):
10% to 75%
Air pressure:
500 hPa to 1060 hPa
AMBIENT CONDITIONS FOR OPERATION
Ambient temperature:
10 °C to 40 °C
Relative humidity (non-condensing):
30% to 75%
Air pressure:
700 hPa to 1060 hPa