![MIMSAL MIMLED 1000 Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/mimsal/mimled-1000/mimled-1000_instructions-for-use-manual_1800503005.webp)
5
MIMLED 600 / MIMLED 1000
Please also refer to the special safety instructions in the following chapters.
2.1 STANDARDS AND GUIDELINES
The device meets the safety requirements the following standards, laws and directives:
•
Medical Device Act (MPG in Germany);
•
EN 60601-1: Medical Electrical Equipment – Part1: General requirements for basic safety and essential
performance.
2.2 SPECIFIC PURPOSE
•
The devices are exclusively used to support and position the
MIMLED 600
and
MIMLED 1000
light
heads as well as supply the devices with power.
•
The devices are suitable for continuous operation.
•
The devices are not suitable for the fi eld of home care
2.3 IMPROPER USE
•
The devices must not be loaded with additional weights in addition to the light body.
2.4 CONTRAINDICATION
•
The support arm system should not be located in the vicinity of strong magnetic fi elds.
•
Only lights of type
MIMLED 600
and
MIMLED 1000
may be con-nected to the support arm system.
GENERAL SAFETY INFORMATION
SYMBOL
DESCRIPTION
RISK OF TIPPING:
The device can take a maximum payload of 2.9 kg. If the maximum payload is exceeded, the
device may tip over, fall onto people and cause serious injury.
Do not climb or lean on the stand. Do not attach other loads to the device.
RELOCATION:
Before moving (arrow 2) device stand, make sure that the spring arm is in
the lowest position (arrow 1).
•
Make sure that the lockable castors are unlocked.
•
Pay attention to bumps, door sills, steps on wheel chair entrances or other obstructions.
•
Make sure you use an appropriate vehicle speed so that it is possible to stop and and avo-
id collision at any time. Pay attention to steep fl ooring.
2
GENERAL SAFETY INFORMATION