Dear Customer,

Thank you for choosing a Milwaukee Instruments 

product. This manual will provide you with the necessary 

information for the correct use of the instrument. Please 

read it carefully before using the meter. 

SPECIFICATIONS

Range

0.00 to 20.00 mg/L Cl

-

Resolution

0.01 mg/L

Precision

±1.0 mg/L @10.0 mg/L

Light Source

Blue LED 466 nm

Light Detector

Silicon Photocell and 466 nm narrow 

band interference filter

Method

Adaptation of mercury (II) thiocyanate 

method

Environment

0 to 50 °C (32 to 122 °F) 

100% RH max

Battery Type

9 volt (1 pc)

Auto-Shut off

After 10’ of non-use

Dimensions

192 x 104 x 52 mm 

(7.5 x 4.1 x 2”)

Weight

380 g

MEASUREMENT PROCEDURE

1.  Turn the meter on by pressing ON/OFF.

2.  When the LCD displays “- - -”, the 

meter is ready.

3.  Fill two cuvets, one with 10 mL of 

distilled water and one with 10 mL of 

the unreacted sample, up to the mark.

4.  Using the 1 mL syringe add 0.5 mL of Cl-1 reagent to 

each cuvet. Replace the caps and swirl gently for 30 

seconds.

 

10 ml

5.  Remove the caps and using the second syringe add 

0.5 mL of Cl-2 reagent to each cuvet. Replace the caps 

and swirl gently for 30 seconds.

 

6.  Place the cuvet with distilled water 

into the holder and ensure that 

the notch on the cap is positioned 

securely into the groove.

7.  Press and hold READ for three seconds and the 

display will show the countdown of the timer. 

Alternatively, wait for 2 minutes.

 

2’

8.  Press ZERO and “SIP” will blink on the 

display.

9.  After a few seconds the display will 

show “-0.0-”. The meter is now zeroed 

and ready for measurement.

 

 

10. Remove the cuvet with distilled water.

11.  Insert the cuvet with the sample and ensure that the 

notch on the cap is positioned securely into the groove.

12. Press READ and “SIP” will blink 

during measurement.

13. The instrument directly displays 

concentration in mg/L of chloride.

Interferences

 

y

pH of the sample should be about 2. For alkaline 

samples, neutralize before adding reagents.

 

y

Color of the sample will cause interferences, they should 

be adequately treated before performing the test.

 

y

Suspended matter in large amounts should be 

removed by prior filtration.

GUIDE TO DISPLAY CODES

This prompt appears for 1 second 

each time the instrument is turned 

on.
The dashes “- - -” indicates that the 

meter is in a ready state and zeroing 

can be performed.
Sampling In Progress. Flashing “SIP” 

prompt appears each time the meter 

is performing a measurement.

“-0.0-”, the meter is in a zeroed state 

and measurement can be performed.

The blinking “BAT” indicates that the 

battery voltage is getting low and the 

battery needs to be replaced.
“-bA-”, the battery is dead and must 

be replaced. Once this indication 

is displayed, the meter will lock up. 

Change the battery and restart.
“Conf”, the meter has lost its 

configuration. Contact your dealer 

or the nearest Milwaukee Customer 

Service Center.

ERROR MESSAGES

On Zero Reading

 

y

Blinking “-0.0-” indicates that the 

zeroing procedure failed due to a 

low signal-to-noise ratio. In this 

case press ZERO again.

 

y

“no L”, the instrument cannot 

adjust the light level. Please check 

that the sample does not contain 

any debris.

 

y

“L Lo”, there is not enough light to 

perform a measurement. Please 

check the preparation of the zero 

cuvet.

 

y

“L Hi”, there is too much light to 

perform a measurement. Please 

check the preparation of the zero 

cuvet.