Automatic temperature 

compensation indicator

Instability 

indicator

Calibration 

indicator

Low battery 

indicator

Measuring 

unit

Main 

LCD

Secondary 

LCD

Temperature 

Unit

OPERATIONAL GUIDE

 

y

Remove the electrode cap and immerse the 
tester in MA9015 storage solution for 2 hours to 
activate the electrode.

 

y

Turn the tester on by pressing the ON/OFF 
button. All the used segments on the LCD will 
be visible for 1 second or as long as the button 
is pressed.

 

y

Immerse the electrode in the solution to be 
tested. Stir gently and wait for the reading to 
stabilize, i.e. the instability indicator (hourglass) 
on the LCD turns off.

 

y

The pH value is automatically compensated for 
temperature and will be displayed on the main 
LCD, while the temperature is shown on the 
secondary LCD.

 

y

To freeze the display

, while in measurement 

mode, press and hold the SET/HOLD button. 
The “HOLD” message appears on the secondary 
display and the reading will be frozen on the 
LCD.
Press any button to return to normal mode.

 

y

To turn the tester off

, press the ON/OFF 

button. The “OFF” message will appear on the 
secondary display. Release the button.

Note:

 

y

Before taking any measurement, make sure the 
tester is calibrated (the CAL tag is on).

 

y

After use always turn the tester OFF, rinse the 
electrode with water to minimize contamination 
and store it with a few drops of storage (MA9015
or pH7 (M10007) solution in the protective cap.

NEVER USE DISTILLED OR DEIONIZED WATER FOR 
STORAGE PURPOSE.

CALIBRATION PROCEDURE

For better accuracy, frequent calibration of 

the tester is recommended. Calibration is also 

necessary after electrode replacement, after 

testing aggressive chemicals and where extreme 

accuracy is required.

 

y

From normal operation mode, press and hold 

the ON/OFF/CAL button until the “OFF” message 

on the secondary LCD is substituted by “CAL”. 

Release the button.

 

y

The instrument enters the calibration mode by 

displaying “pH 7.01 USE” (or “pH 6.86 USE” if the 

NIST buffer set was selected).

 

y

For a 

single-point calibration

, immerse the 

electrode in any buffer, i.e. pH 4.01, 7.01 (or 6.86), 

10.01 (or 9.18).

 

y

The tester activates the automatic buffer 

recognition. If no valid buffer is detected, the 

tester keeps the USE indication active for 12 

seconds, and then replaces it with WRNG, 

indicating that the sample being measured is 

not a valid buffer. Otherwise, if a valid buffer is 

detected, then its value is shown on the primary 

display, and REC appears on the secondary LCD.

 

y

If the pH 7.01 (or pH 6.86) was used, press the 

SET button to exit the Calibration mode and the 

“OK 1” message will appear on the display. The 

first calibration point is stored and the tester 

returns to normal measurement mode.

For better accuracy, it is always recommended 

to perform a 2-point calibration.

 

y

For a 

two-point calibration

, immerse the 

electrode in pH 7.01 (or pH 6.86) buffer solution.

 

y

After the first point has been accepted, the 

tester will then ask for the second buffer and 

the message “pH 4.01 USE” will appear.

 

y

Rinse the electrode and immerse it in the 

second solution (pH 4.01, 10.01 or 9.18).

 

y

If a valid buffer value is detected, the REC 

message is displayed and the tester completes 

the calibration procedure. The LCD shows the 

accepted value with the “OK 2” message and the 

instrument returns to the normal measurement 

mode. Otherwise, if no valid buffer is detected, 

the tester displays the WRNG message.

Note: When the calibration procedure is 

completed, the CAL tag is turned on.

 

y

To quit the procedure

 and return to the last 

calibration data, after entering the calibration 

mode press the ON/OFF button. The secondary 

LCD displays “ESC” for 1 second and then the 

tester returns to the normal measurement 

mode.

 

y

To reset to the default values

 and clear a 

previous calibration, press the SET/HOLD button 

after entering the calibration mode and before 

the first point is accepted. The secondary LCD 

displays “CLR” for 1 second, the tester resets to 

the default calibration and the CAL tag on the 

LCD turns off.

SETUP

Setup mode allows the selection of temperature 

(°C or °F) and pH buffer set for calibration.

To enter the Setup mode, press the ON/OFF 

button until “CAL” on the secondary LCD is 

replaced by “TEMP” and the current temperature 

unit (e.g. TEMP °C). Then:

 

y

for °C/°F selection:

 use the SET/HOLD button; 

then press the ON/OFF button once to enter 

the buffer set selection or twice to return to the 

normal measurement mode.

 

y

to change the calibration buffer set:

 after 

setting the temperature unit, press ON/OFF 

once and select the buffer set (“pH 7.01 BUFF” or 

“pH 6.86 BUFF” for NIST) by using the SET/HOLD 

button. Press ON/OFF to return to the normal 

measurement mode.

ELECTRODE REPLACEMENT

 

y

Remove the protective cap and unscrew the 

plastic ring on the top of the electrode.

 

y

Pull out the MI56P electrode and replace it with 

a new one.

 

y

Make sure the gaskets are in place before 

screwing back the ring.

BATTERY REPLACEMENT

When the batteries become weak, the battery 

symbol on the LCD will light up to advise that only a 

few hours of working time is remaining.

The tester is also provided with BEPS (Battery Error 

Prevention System), which avoids any erroneous 

readings due to low battery level by automatically 

switching the tester off.

It is recommended to replace the batteries 

immediately.