32
MW605 Dissolved Oxygen & Temperature Portable Meter
Delete All
1. Press RCL to access the logged data.
LCD displays “PLEASE WAIT” followed by “LOG RECALL” with
ACCEPT tag blinking and the number of stored logs.
2. Press LOG/CLEAR to delete all logs.
“CLEAR ALL” is displayed with ACCEPT tag blinking.
3. Press GLP/ACCEPT to confirm (to exit, press ESC or CAL/EDIT;
or LOG/CLEAR).
“PLEASE WAIT” is displayed with a percentage counter, until
all logs are deleted. After deletion “CLEAR DONE” message
displays briefly. Display returns to the log recall screen.
11.2.3. Exporting Data
PC Export
1. With the meter on, use the supplied micro USB cable to
connect to a PC.
2. Press SETUP then CAL/EDIT.
3. Use the
keys and select “EXPORT TO PC”.
The meter is detected as a removable drive. LCD displays the
PC icon.
4. Use a file manager to view or copy files on the meter.